Instruction Manual AquaMaster
Mounting and installation
30
12745E/4
4.6
Fasten photometer to base plate
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Fasten the photometer mounting bracket to
the base plate. Make sure that the two posi-
tioning pins (circles) are inserted into the holes
on the mounting bracket of the photometer.
Ensure that the outlet tube of the in-
strument is aligned to the drain funnel (the
union piece of the drain funnel may have to
be loosened and the funnel aligned).
2.
Remove the optics unit of the photometer ac-
cording to the Instruction Manual.
3.
Align the photometer according to the Instruc-
tion Manual.
4.
4.1: Insert the angled inlet tube from above
into the union piece.