background image

Signet 2820 Series Conductivity Sensor Instruction Manual

1.  Wiring

2.  Recommended Position

3.  2819/2820/2821 In-line Installation

SAFETY INSTRUCTIONS  FOR IN-LINE ELECTRODE INSTALLATION

1. 

Do not remove from pressurized lines.

2. 

Do not exceed maximum temperature/pressure specifi cations.

3. 

Wear safety goggles or face shield during installation/service.

4. 

Do not alter product construction.

5. 

Disconnect instrument power before wiring this sensor. 

Failure to follow safety instructions may result in severe personal injury!

Customer
supplied 
pipe tee/
reducer

Standard
fitting kit

Hole up

Mark hole 
position 

3/4 in. NPT

Hand tighten only!

Optional
fitting kit

Hole up

Mark hole 
position 

Customer
supplied
pipe tee/
reducer

1/2 in. NPT

Hand tighten only!

O-ring

O-ring

Sealant

Sealant

Signet 5800CR 

•  Use three conductor shielded cable for cable extensions up to 30 m (100 ft) max.
•  Shield must be maintained through cable splice.
•  Disable power to instrument before connecting this sensor.

RED

WHITE

BLACK

SILVER (SHLD)

Shld

Signal IN

T

e

mp. IN

Iso. Gnd

CH 2

CH 1

RED

SILVER (SHLD)

BLACK

SILVER (SHLD)

BLACK

C

NO

Tx   GND  Rx 

CH2

S2    T2

S1   T1   SG 

C

NO

R+

R-

RLY3

RLY4

SERIAL

ANL2

CH1

1       2       3         4       5         6       7         8       9       10  11  12   13      14      

WHITE

RED

WHITE

Signet 9050CR

In-line

2821-1

Use caution to avoid air bubble 
or sediment trapping inside the 
electrode cavity.

Submersible

OR

*3-2820.090*

3-2820.090-1 Rev. 

11/04  English

English

Warning!

Summary of Contents for 2820-1

Page 1: ...position 3 4 in NPT Hand tighten only Optional fitting kit Hole up Mark hole position Customer supplied pipe tee reducer 1 2 in NPT Hand tighten only O ring O ring Sealant Sealant Signet 5800CR Use three conductor shielded cable for cable extensions up to 30 m 100 ft max Shield must be maintained through cable splice Disable power to instrument before connecting this sensor RED WHITE BLACK SILVER ...

Page 2: ...onal fitting kit 3 4 in NPT 1 2 in NPT 1 Customer supplied pipe Customer supplied coupling Hand tighten only Standard fitting kit Optional fitting kit O ring Sealant Sealant Optional submersion fitting kit 3 2820 390 1 2 1 1 4 turns past finger tight 3 4 in NPT Customer supplied pipe Customer supplied coupling Sealant Sealant 8 2823 Submersible Installation Fill with 3 to 4 in of sealant Attach 3 ...

Page 3: ... S1 2820 S1 2821 S1 2819 T1 2820 T1 2821 T1 2819 X1 cell 0 01 2819 X1 range 0 010 to 100 µS 10 kΩ to 100 MΩ 2820 X1 cell 0 1 2820 X1 range 1 to 1000 µS 2821 X1 cell 1 0 2821 X1 range 10 to 10 000 µS Sanitary fitting size 1 and 11 2 in Temperature compensation PT1000 Wetted materials O rings EPR Insulator material PTFE Electrodes 316 stainless steel or titanium Sanitary fitting 316 stainless steel or...

Page 4: ... 6 9 bar 100 psi Max temperature 95 C 203 F Quality Standard CE 2823 1 Cell 20 0 Range 200 to 400 000 µS Temperature compensation PT1000 Wetted materials O rings EPR Insulator material PTFE Electrodes 316 stainless steel Fitting 316 stainless steel Max pressure 6 9 bar 100 psi Max temperature 150 C 302 F Quality Standard CE 8 0 in 203 mm 15 ft 4 6 m cable standard 3 4 in NPT Reversible fitting ass...

Reviews: