Signet 2507 Mini Flow Sensor

3

5. Calibration

The K-Factors listed here represent the number of pulses the sensor will generate for each measured engineering unit. They are 
listed in Liters and in U.S. gallons for each sensor model. 

Sensor

Model

Code

Flow

Insert

K-FACTORS

Pulses per

LITER

Pulses per

U.S. Gallon

3-2507.100-1V

198 801 731

1 mm

3413

12918

3-2507.100-2V

198 801 732

2 mm

1687

6385

3-2507.100-3V

198 801 733

3 mm

1045

3955

3-2507.100-4V

198 801 734

4 mm

721

2729

3-2507.100-6V

198 801 736

NONE

382

1446

FLOW 

Red 

Black 

Shield 

Signet 
5500/5600 

 2507 

Freq. IN

Sen. Pwr

Freq. IN 

Iso. Gnd

Open Collector  
Sensor 

Std. Sensor 

Red

Black

Shield

Signet 8550-1 

FLOW 

 2507 

Sensr Gnd 

(SHIELD) 

 

Sensr IN 

(RED) 

 

Sensr V+ 

(BLACK) 

Signet 8900 

FLOW 

FLOW

 2507

Input 

Gnd. 

10k Ω 

+

5 to 24  

VDC 

-

Black   

Shield 

Red 

•  Pull-up resistor required (10 kΩ recommended).  Consult your instrument manual for  
 additional 

information. 

Other 

instrument 

Other Brands 

External Power supply 

3-8900.621C 

I/O Module 3-8900.401-X 

10 

11 

12 

13 

14 

+5VDC (Black) 

Freq. Input (Red) 

GND (Shield) 

+5VDC (Black) 

Freq. Input 2 (Red) 

S  L (Red) 

GND (White/Shield) 

+5VDC (Black) 

S  L (Red) 

GND (White/Shield) 

 

Analog Output 1 

 

Analog Output 2 

(if applicable) 

(if applicable) 

Frequency 

Input 

Frequency 

Input 2 

OR 

S3L 

Input 

 

S3L 

Input 

2507 

 2507

FLOW

Instrument Connections

Assemble

 (counterclockwise)

Disassemble 

(clockwise)

Upper body

Lower Body

Assembly
Tabs (3)

FLOW

Inlet port

Outlet port

flow 

insert

Rotor

Quad ring seal

WARNING!

Do not use tools of any kind on the sensor body or port 
connections. Hand tighten only! Excessive force will 
damage sensor.

6.  Replacing The Flow Insert

Sensor range can be modifi ed by changing the fl ow insert. The 
sensor must be removed from service and disassembled prior to 
installing the new fl ow insert. See Specifi cations (Section 2) for 
fl ow range data.

Flow Insert Replacement Procedure:
1.  Depressurize system and remove sensor.
2.  Rotate the upper sensor body clockwise until it releases from 

the lower half, then lift off.

3.  Remove rotor and quad ring seal from lower body.
4. Push 

the 

fl ow insert outward using a small screwdriver.

5.  Install the new fl ow insert (small diameter inward) with the 

eraser end of a pencil. Apply light pressure until insert seats 
against the step in the lower body. 

Do not force!

6.  Install rotor into lower body. Spin rotor with fi nger and check 

for free rotation. If rotor hits fl ow insert, remove rotor and 
push insert back until free rotor rotation is established. Use a 
rounded object like a pen or pencil body to adjust fl ow insert 
depth.

7.  Install rotor, quad ring, and upper body. 

Hand tighten only! 

Do not overtighten upper body or the lower body assembly 
tabs will break. 

8.  Reprogram instrument with new K-Factor, see Calibration 

(Section 5).

IMPORTANT!

• 

K-Factors must be considered as approximate 
values.

• 

The number of pulses per volumetric unit 
may vary depending on the medium and the 
installation.

• 

For optimum performance, the system must 
be calibrated after installation.

Summary of Contents for 2507

Page 1: ...personal injury 1 Description The Signet 2507 Mini Flow Sensor contains a free running rotor which is driven by the fluid flow Within the given measurement range the rotational speed of the rotor is proportional to the fluid flow rate Permanent magnets built into the rotor actuate an electronic switch in the top of the sensor generating a square wave output signal proportional to flow rate Both op...

Page 2: ... 4 in straight threads to 1 4 in NPT pipe threads Hand tighten only Apply 1 2 turns of PTFE sealing tape to all threaded connections to prevent leaks Installation Hints Avoid vibrations and shocks Avoid solids in the fluid Install a filter or line strainer upstream to protect sensor 1 2 Sensor Ground silver wire Sensor Power 5 to 24 VDC black wire Sensor Signal Output Red wire Max cable length 300...

Page 3: ...Assembly Tabs 3 FLOW Inlet port Outlet port flow insert Rotor Quad ring seal WARNING Do not use tools of any kind on the sensor body or port connections Hand tighten only Excessive force will damage sensor 6 Replacing The Flow Insert Sensor range can be modified by changing the flow insert The sensor must be removed from service and disassembled prior to installing the new flow insert See Specific...

Page 4: ... Rotor 2507 3 2507 080 3 198 801 547 Quad Ring 2507 3 2507 080 5 198 801 508 DIN Connector 2507 3 2507 081 1 198 801 548 1 mm Insert available in Europe only 3 2507 081 2 198 801 502 2 mm Insert 3 2507 081 3 198 801 503 3 mm Insert 3 2507 081 4 198 801 558 4 mm Insert 5523 0222 159 000 392 Cable per foot 2 cond w shield 22 AWG Ordering Information Sensor Part Number 3 2507 100 Mini flow low flow s...

Reviews: