![Sigma 6-16KHS Operating Manual Download Page 66](http://html1.mh-extra.com/html/sigma/6-16khs/6-16khs_operating-manual_1267909066.webp)
Refrigerated Centrifuge with Heating System Sigma 6-16KHS
8 Maintenance and service
66 / 93
Version 10/2012, Rev. 1.19 of 21/03/2019
• sb
Translation of the original operating manual, part no. 0702302
8.1.4
Load bearing bolts
Only greased load-bearing bolts ensure a uniform swing-out of the buckets
and, therefore, the smooth operation of the centrifuge. Load-bearing bolts
that are insufficiently greased may cause the centrifuge to stop due to an
imbalance.
• Clean the load-bearing bolts and bucket groove in order to remove the
old grease.
• Apply a small amount of heavy-duty grease for load-bearing bolts (ref.
no. 71401, see the following picture) to both load-bearing bolts of a
bucket.
Fig. 33: Sufficient quantity of grease for one bolt
• Install the bucket and swing it manually back and forth once in order to
distribute the grease.
• Repeat this process with all the other buckets.