Laboratory Centrifuge Sigma 3-16L
Laboratory Centrifuge Sigma 3-16L IVD
6 Using the centrifuge
36 / 79
Version 03/2014, Rev. 2.11 of 11/08/2017
• sb
Translation of the original operating manual, part no. 0700902
Automatic rotor identification system
1
If the centrifuge is equipped with an automatic rotor identification system,
the input mode will be activated automatically if the system detects a
different rotor with several different types of possible buckets than the rotor
that is set. The bucket that is displayed is the bucket with the lowest
maximum speed. You can only select different types of buckets for the
identified rotor. If the input mode is aborted, this bucket will be stored
nonetheless.
This prevents the maximum permissible speed from being exceeded.
Pos : 134 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 135 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S/ Spinc ontrol L (Z ent6) /060- 0030-0020- 0091 Besc hleunigungs- und Brems kur ven - ohne Temp @ 27\ mod_1405319282549_68.doc x @ 192843 @ 4 @ 1
6.3.2.8
Acceleration and deceleration curves
Acceleration
This function is used to select an acceleration curve. The system offers 10
fixed, programmed acceleration curves (curves 0-9).
Brake
This function is used in order to select a curve that decelerates the
centrifuge to a standstill. Deceleration curves are inverted images of the
acceleration curves and are labelled with identical numbers. Deceleration
curve no. 0 represents a brakeless deceleration.
Fig. 17: Preselection of a curve; here: preselection of an acceleration curve
Pos : 136 /010 Uni vers almodul e/ Seit enwechs el @ 0\ mod_1202116244312_0.doc x @ 105 @ @ 1
1
not available for Sigma 4-5L