- 19 -
e.
To prevent mistaking into the calibration setting interface block, before entering the
interface, should enter the password authentication interface, continuous inputting
"STORE" button four times (password “□”box fill four small black spots, more or less
filling are invalid), and press the confirm button "OK" to enter the settings screen, press
the other keys are invalid during this process.
5.3.4.6 Selection frequency and probe
a.
Press upward keys” ” or downward key " " to select the project (tick inside the
block), press "OK" key to save and return to “remind changing probe, re-calibration”
interface, seeing the picture below. Then press measurement key “MEAS” or wait the
interface timing finished automatically to go into the measurement interface.
b.
60 kHz and 500 kHz is the working frequency of the instrument probe, 60 kHz
φ14mm is for aviation industry standard, and 500 kHz φ 8mm is for detection of sheet
material or small area of the work pieces material.
c.
2008B1 Sigma type conductivity meter can not enter the interface. Sigma 2008B type
machine can be selected according to the working frequency of the probe.