Refrigerated Centrifuge Sigma 2-16KL
8 Maintenance and service
56 / 82
Version 08/2012, Rev. 1.24 of 28/04/2022
• sb
Translation of the original operating manual, part no. 0700702
Pos : 188 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/080 War tung und I nst andhaltung/ 080-0010- 0040 Tragbolz en (nur 2- 6, 2-6E, 2- 7, 2-16P, 2-16KL) @ 44\m od_1467889556212_68.docx @ 436524 @ 3 @ 1
→
8.1.5
Load-bearing bolts
Plastic bucket
NOTE
The load-bearing bolts of the rotor must not be greased if plastic buckets
are used!
Aluminium buckets
Only greased load-bearing bolts ensure a uniform swing-out of the buckets
and, therefore, the smooth operation of the centrifuge. Load-bearing bolts
that are insufficiently greased may cause the centrifuge to stop due to an
imbalance.
• Clean the load-bearing bolts and bucket groove in order to remove the
old grease.
• Apply a small amount of heavy-duty grease for load-bearing bolts (ref.
no. 71401, see the following picture) to both load-bearing bolts of a
bucket.
Fig. 29: Sufficient quantity of grease for one bolt
• Install the bucket and swing it manually back and forth once in order to
distribute the grease.
• Repeat this process with all the other buckets.