SIGLENT Digital Multimeter SC1016 User’s Guide
WWW.SIGLENT.COM
SIGLENT Digital Multimeter SC1016 User’s Guide
Safety
SIGLENT thanks you for purchasing the SC1016 module. To achieve the best performance from the product, please read this
guide carefully. To avoid electrical shock and personal injury, please don’t use the product to measure signals that exceed the
published specification.
Specifications
Max AC Voltage
125V rms or 175V peak, 100kHz,
0.3 A switched, 125VA (resistive load)
Maximum DC Voltage
110V, 1A switched, 30VDC (resistive load)
Contact Life
>
100000 operations
,
at 1A 30VDC(at 0.5 Hz)
>
100000 operations
,
at 0.3A 125VDC (at 0.5 Hz)
Contact Resistance
75 mΩ
(
maximum at 6VDC, 1A
)
Actuation Time
5ms maximum on/off
M a x i m u m s w i t c h i n g
voltage
250 VAC, 220 VDC
Maximum switching power
62.5VA / 30W
Insulation Resistance
Minimum 1G ohm
(
500VDC
)
Connector Type
Clamp terminal, #24 AWG wire size
Prenotion: To avoid breaking the product, please do not pull it away from multimeter when measurement is executed.
Channel Capabilities
Item
No. of wires
No. of channels
DCV, ACV
[1]
2 wires (H, L)
12 (CH1~CH12)
DCI, ACI
[2]
2 wires (H, L)
4 (CH13~CH16)
(2A Range)
2W Resistance
2 wires (H, L)
12 (CH1~CH12)
4W Resistance
4 wires (Input H, L
+ sense H, L)
6 pairs (CH1 [input]
& CH7 [sense], 2&8, •••, 6&12)
Capacitance
2 wires (H, L)
12 (CH1~ CH12)
Diode/Continuity
2 wires (H, L)
12 (CH1~ CH12)
Period/Frequency
2 wires (H, L)
12 (CH1~ CH12)
Temp(Thermocouple)
Temp (RTD)
2 wires (H, L)
2 wires (H, L)
12 (CH1~ CH12)
12 (CH1~ CH12)
Remarks:
[1]Voltage range : <125VAC, 110VDC
[2] For continuous current < 2.2A, Accuracy ± (% 3 (reading) + 0.02% (range)).
Summary of Contents for SC1016
Page 1: ...SC1016 User s Guide...