SNA5000A Vector Network Analyzer User Manual 68
option, and select the corresponding calibration method from the Cal Type drop-down option,
as shown in the figure below:
Figure 3-18 Calibration wizard interface
After the calibration is set, a specific calibration item interface will appear as shown in the
figure below. Here, taking the two-port SOLT calibration method as an example, it includes
several calibration items: OPEN, SHORT, LOAD, THRU, ISOLATION, among which the
ISOLATION calibration item may not be calibrated. There is no sequence relation between
these calibration terms. Simply select the corresponding calibration kit and click the
corresponding calibration item menu on the interface.
After completion of the corresponding calibration item, the percentage of calibration
progress will be displayed in the lower-left corner of the interface. When all the calibration
steps are completed, the calibration progress bar will show 100%. Click Finish to exit the
calibration interface, calibration results are automatically loaded in the current screen.
Figure 3-19 Calibration wizard interface
To Save the current calibration data for future use, press the
Save Recall
button and use
the knob or arrow keys to bring the focus to the
Save State
menu bar. Select
State+Cal Data
from the
Save Type
drop-down option, click
Save State As
and enter the file name, and click
OK
to Save the current calibration Data and the current setting State.