Voltage of the system does not exceed 1000 V.

Make the measuring circuit connections mechanically strong and secure so  
that  they  do  not  accidently  open.  The  conductor  cross  sections  and 
connection points should be designed to avoid excessive heating.  

On  the  300mA  and  10A  ranges,  an  intermittent  sound  signal  warns  you,   
when the measured value exceeds the upper range limit.

The current measuring ranges upto 300 mA are protected to a short circuit  
current of 25 A by a fuse FF 1.6/1000V in conjunction with power diodes. 
The  cutout capacity of the fuse is 10 kA at a rated voltage of 1000 V and 
ohmic load.

The  current  measuring  ranges  up  to  10A  are  protected  by  a  16A/1000  V   
fuse.  The cut-out capacity of the fuse is 30kA at a nominal voltage of 1000  
V and  ohmic load.

A blown fuse is signaled on the LCD the instant a measured quantity having  
a  voltage  of  more  than  4V  is  applied  to  the  corresponding  connection   
sockets.  Then, the digital display (9) shows the word “FUSE”.

After a fuse has blown, eliminate the cause of the overload before using the  
meter again!

Replacement of the fuses is described in section “18. Maintenance”.

AC current measurement with (clip-on) current transformers

See appendix section 20.1.

9. Resistance measurement and continuity test

î

Verify  that  the  device  under  test  is  electrically  dead.  External  voltages   

would falsify the measured result !

î

Set the function selector switch (6) to “ 

”.

î

Connect the device under test as  shown.

î

Zero  adjustment  on  the  measuring  ranges  300 

W

  to  3  k

W

When  measuring  small  resistance  values  on  the  300

W

  and  3k

range,  you 

can  eliminate  the  resistance  of  the  leads  and  transient  resistances  by  zero 
adjustment:

î

Connect the test leads to the meter and join the free ends.

î

Briefly press the yellow multi-function pushbutton (5).

The  meter  acknowledges  zero  adjustment  by  sound  signal,  the  LCD 
shows  “000,00”  (+1  digit)  and  the  symbol  “ZERO”  (18)  is  displayed.  The 
resistance  measured  the  instant  the  pushbutton  is  pressed  is  used  as 
reference value (max. 2000 digits). It is automatically deducted from the values 
measured thereafter. 
You can clear the zero adjustment

-   by  pressing  the  yellow  multifunction  pushbutton  (5)  for  a  long  time, 

clearance is acknowledged by the buzzer sounding twice,

-   by switching the instrument off.

Continuity test with buzzer

With “buzzer” function activated, the meter issues a continuous sound  signal 
below 100.00 on the 300.00 range only.

í

To switch buzzer ON:

Select range 30k

W

, 300k

W

, 3M

W

, or 30M

W

, through manual mode and  briefly 

press the yellow multi function key(5). The meter will show “cont.n”  on display. 
Press “AUTO/MAN” key. The meter will show “cont.y” on display. 
Now press the yellow key(5) again. The symbol      (20) appears on the  display 
when 300.00

range is selected.

í

To switch buzzer OFF:

Select  range  30k

W

,  300k

W

,  3M

W

,  or  30M

W

  ,  through  manual  mode  and   

briefly  press  the  yellow  multi  function  key(5). The  meter  will  show  “cont.y”   
on display. Press “AUTO/MAN” key. The meter will show “cont.n” on display. 
Now press the yellow key(5) again. The symbol      (20) disappears on the 
display when 300.00

 range is selected.

Summary of Contents for Gamma 10

Page 1: ... 2013 P R E C I S I O N I N S T R U M E N T A T I O N Manufactured by Sifam Tinsley Instrumentation Inc 3105 Creekside Village Dr Unit 801 Kennesaw GA 30144 Distributed by Veracious Canada Inc 1101 Britannia Rd E Mississauga ON L4W 3X1 Contact 1 416 640 5364 Fax 1 416 981 3933 E mail info veraciousinc ca Web www veraciousinc com ...

Page 2: ...ased along with this card and relevant cash memo Invoice This warranty entitles you to bring the meter accessories at your cost to the nearest stockiest dealer and collect it afterrepairs NOTRANSPORTATIONCHARGESWILLBEREIMBURSED Thewarrantyisnotvalidinfollowingcases 1 Warranty card duly signed stamped and original cash memo invoice are notsentalongwiththemeter accessories 2 Completewarrantycardisno...

Page 3: ...s and safety precautions 4 2 Switching the meter on 5 3 Function and range selection 6 3 1 Autoranging 6 3 2 Manual range selection 6 4 Liquid crystal display 7 4 1 Digital display 7 4 2 Analog indication 7 4 3 Backlit 7 5 DATA hold compare facility 7 6 Minimum value and maximum value MIN MAX storage facility with time display 8 7 Voltage measurement 8 7 1 Voltage measurement on electrical systems...

Page 4: ...fe and proper condition of the meters and to ensure their safe operation it is absolutely necessary to carefully and completely read these operating instructions before using any meter These instruction must be followed in all respects For your safety and for protection of the meter the Gamma 10 meter is fitted with an automatic terminal blocking system It is coupled with the function selector swi...

Page 5: ...he cause of the overload before using the meter again Meaning of the symbols on the device Warning of a danger point Attention refer to documentation Ground connector Double or reinforced insulation CAT II III IV Instrument for a overvoltage catagory II III or IV EU conformity mark For US Canada Repair replacement of parts and calibration When opening the meter live parts may be exposed Therefore ...

Page 6: ......

Page 7: ...divisions so that variations of the measured values around zero can be observed exactly When the measured value exceeds the range of indication the left triangle 16 is shown before the polarity of the analog indicator switches over after approximately 0 7s Exceeding of the measuring range 30999 digits on the range F 3099 is shown by the right triangle 13 The graduation of the analogue scale is aut...

Page 8: ...tion you can hold the minimum and the maximum measured value which was applied to the input of the meter after activating MIN MAX The most important application is the determination of the minimum and the maximum value for long term monitoring of measured quantities The function MIN MAX can be activated in all ranges except for the diode test However for frequency and capacitance measurement the r...

Page 9: ...tant the pushbutton is pressed is used as reference value max 2000 digits It is automatically deducted from the values measured thereafter you can clear the zero adjustment by pressing the yellow multi function pushbutton 5 for a long time clearance is acknowledged by the buzzer sounding twice by switching the instrument off 7 1 Voltage measurement on electrical systems up to 1000 V with the KS30 ...

Page 10: ... dB V V V mA 0 C A Voltage Measurement on electrical system up to 1000 V with the KS30 measuring adapter 1 min continuou sly 1 min 7 Range 3V 30V 300V U 1V 10V e eff rms eff rms U 10V 100V e eff rms eff rms U 100V 1000V e eff rms eff rms 8 Without sensor ...

Page 11: ...power systems when the current circuit is protected by a fuse or a circuit breaker of 20 A and when the nominal Note The multimeter does not contain termination resistors The input impedance is 5 MΩ respectively In order to measure correctly on unterminated objects you have to connect the termination resistor to the terminals Please take the power dissipated by the termination resistor into consid...

Page 12: ... valid for 3 of measuring range 6 1 30 6 1 30 6 1 30 6 1 30 6 1 30 6 1 5 30 6 1 5 30 6 1 5 30 6 1 5 30 6 1 5 50 6 1 5 30 A 20 1 1 mA 20 1 1 V 20 1 2 Transformer output F W Hz dB V V V mA 0 C A F W Hz dB V V V mA 0 C A F W Hz dB V V V mA 0 C A ...

Page 13: ...nder test as shown î Zero adjustment on the measuring ranges 300 W to 3 kW When measuring small resistance values on the 300W and 3kW range you can eliminate the resistance of the leads and transient resistances by zero adjustment î Connect the test leads to the meter and join the free ends î Briefly press the yellow multi function pushbutton 5 The meter acknowledges zero adjustment by sound signa...

Page 14: ...les monitor can also be used for the DOS version w A hard disk with 3 MB free storage space w A 3 5 disk driver fas disks with 1 4 MB storage capacity w A MICROSOFT compatible mouse If you wish to make hard copies w an EPSON compatible printer when using the DOS version w a printer supported by WINDOWS when using WINDOWS A F W Hz dB V V V mA 0 C 00000 A F W Hz dB V V V mA 0 C 0 6000 A F W Hz dB V ...

Page 15: ...the device under test is electrically dead External voltages would falsifythemeasuredresult ê Setthefunctionselectorswitch 6 to ê Connectthedeviceundertestasshown Forwarddirectionand orshortcircuit The meter displays the forward voltage in Volts As long as the voltage drop does not exceed the maximum display value of 3 0999 V you can also test several series connectedelementsorreferencediodeswiths...

Page 16: ...nt w bypressingtheyellowmulti functionpushbutton 5 foralongtime Themeter acknowledges this with the buzzer sounding twice The voltage measuring rangelastselectedismaintained w byoperatingthefunctionselectorswitch 6 13 Temperaturemeasurement With Pt 100 and Pt 1000 temperature sensors you can measure temperatures on 0 0 therangefrom 200 100 C 850 C 0 ê Setthefunctionswitch 6 to C ê Connectthesensor...

Reviews: