Chapter 4 Serial Communications
Sierra 240/241 Instruction Manual
4-10
IM-240
Fast Key Sequence
Use password 16363.
Sequence
Description
Access
Notes
1,1,1
Snsr
View
Primary variable value
1,1,2
AI % Rnge
View
Analog output % range
1,1,3
AO1
View
Analog output, mA
1,2,1
Test Device
N/A
Not used
1,2,2,1
4 mA
View
Loop test, fix analog output at 4 mA
1,2,2,2
20 mA
View
Loop test, fix analog output at 20 mA
1,2,2,3
Other
Edit
Loop test, fix analog output at mA value entered
1,2,2,4
End
Exit loop test
1,2,3,1,1
4 mA
N/A
Not used, apply values
1,2,3,1,2
20 mA
N/A
Not used, apply values
1,2,3,1,3
Exit
Exit apply values
1,2,3,2,1
PV LRV
Edit
Primary variable lower range value
1,2,3,2,2
PV URV
Edit
Primary variable upper range value
1,2,3,2,3
PV USL
View
Primary variable upper sensor limit
1,2,3,2,4
PV LSL
View
Primary variable lower sensor limit
1,2,4
D/A Trim
Edit
Calibrate electronics 4mA and 20mAvalues
1,3,1
Tag
Edit
Tag
1,3,2
PV unit
Edit
Primary variable units
1,3,3,1
PV LRV
Edit
Primary variable lower range value
1,3,3,2
PV URV
Edit
Primary variable upper range value
1,3,3,3
PV LSL
View
Primary variable upper sensor limit
1,3,3,4
PV USL
View
Primary variable lower sensor limit
1,3,4,1
Distributor
N/A
Not used
1,3,4,2
Model
N/A
Not used
1,3,4,3
Dev id
View
Device identification
1,3,4,4
Tag
Edit
Tag
1,3,4,5
Date
Edit
Date
1,3,4,6
Write Protect
View
Write protect
1,3,4,7
Descriptor
Edit
Vortex flowmeter
1,3,4,8
Message
Edit
32 character alphanumeric message
1,3,4,9
PV snsr s/n
View
Primary variable sensor serial number
1,3,4,menu
Final assy #
Edit
Final assembly number
1,3,4,menu,1
Universal Rev
View
Universal revision
1,3,4,menu,2
Fld dev Rev
View
Field device revision
1,3,4,menu,3
Software Rev
View
Software revision
1,3,5
PV Xfer fnctn
View
Linear
1,3,6
PV Damp
Edit
Primary variable damping (time constant) in seconds
1,4,1,1
PV
View
Primary variable value
1,4,1,2
PV Sensor Unit
Edit
Primary variable units
1,4,1,3
Sensor Information
View
PV LSL, PV USL, PV Min span
1,4,2,1
Snsr Damp
Edit
Primary variable damping (time constant) in seconds
1,4,2,2,1
PV LRV
Edit
Primary variable low range value
1,4,2,2,2
PV URV
Edit
Primary variable upper range value
1,4,2,3,1
PV LRV
Edit
Primary variable low range value
1,4,2,3,2
PV URV
Edit
Primary variable upper range value
1,4,2,4
Xfer Fnctn
View
Linear
1,4,2,5
AI % rnge
View
Analog output % range
1,4,3,1,1
AO1
View
Analog output, mA
1,4,3,1,2
AO alarm typ
N/A
Not used
Continued on next page
Summary of Contents for INNOVA-MASS 240-VT
Page 124: ...Appendix A Specifications Sierra 240 241 Series Instruction Manual A 2 IM 240...
Page 125: ...Sierra 240 241 Series Instruction Manual Appendix A Specifications IM 240 A 3...
Page 126: ...Appendix A Specifications Sierra 240 241 Series Instruction Manual A 4 IM 240...
Page 127: ...Sierra 240 241 Series Instruction Manual Appendix A Specifications IM 240 A 5...
Page 128: ...Appendix A Specifications Sierra 240 241 Series Instruction Manual A 6 IM 240...
Page 129: ...Sierra 240 241 Series Instruction Manual Appendix A Specifications IM 240 A 7...
Page 130: ...Appendix A Specifications Sierra 240 241 Series Instruction Manual A 8 IM 240...
Page 131: ...Sierra 240 241 Series Instruction Manual Appendix A Specifications IM 240 A 9...
Page 132: ...Appendix A Specifications Sierra 240 241 Series Instruction Manual A 10 IM 240...
Page 133: ...Sierra 240 241 Series Instruction Manual Appendix A Specifications IM 240 A 11...
Page 134: ...Appendix A Specifications Sierra 240 241 Series Instruction Manual A 12 IM 240...
Page 135: ...Sierra 240 241 Series Instruction Manual Appendix A Specifications IM 240 A 13...