Sierra Instruments Instruction Manual
Smart-Trak
®
2 Series 100
64
Span:
This allows the user to enter an offset to the span to adjust the meter outputs to
match another device or field conditions. The span is a MULITPLIER and will
multiply all outputs by the entered amount.
For example, the meter reads 98. The desired reading is 100. A span value of 1.02
will adjust the meter output to read 100.
Caution
: this number should not be adjusted beyond 0.8 to 1.2.
Change full scale:
This setting allows you to change the full scale of your instrument