448
Specials
Radiation exposure to patients
• CARE Vision applies continuous exposure at moder-
ate mA levels.
• Due to the potentially long scan times and the lim-
ited scan volume, the dose for certain slices may
increase to levels significantly higher than those
known from standard CT applications.
• The patient’s exposure levels is usually estimated by
CTDIvol.
• This unit is designed to give the average dose in the
scanned volume.
• Before starting the scan, the dose rate (CTDIvol in
mGy per second) is displayed on the monitor.
• During the CARE Vision scan, the accumulated dose
(CTDI vol) reflecting the patient exposure is dis-
played on the monitor. The display scale ranges from
zero to 3000 mGy.
• If the table is shifted during the examination, the
accumulated dose will be distributed to different
slices and will be lower than indicated by the display.
• When a new scan is loaded, the dose display starts
again from zero.
Summary of Contents for SOMATOM
Page 4: ...4 Overview Respiratory Gating 484 Children 508 ...
Page 5: ...Overview 5 ...
Page 121: ...Workflow Information 121 ...
Page 135: ...Contrast Medium 135 ...
Page 157: ...Application Information 157 Program to define capture options ...
Page 161: ...Application Information 161 ...
Page 167: ...Head 167 ...
Page 183: ...Head 183 ...
Page 187: ...Head 187 ...
Page 191: ...Head 191 ...
Page 195: ...Head 195 ...
Page 211: ...Neck 211 ...
Page 217: ...Neck 217 ...
Page 223: ...Shoulder 223 ...
Page 239: ...Thorax 239 ...
Page 249: ...Thorax 249 ...
Page 269: ...Abdomen 269 ...
Page 273: ...Abdomen 273 ...
Page 277: ...Abdomen 277 ...
Page 287: ...Abdomen 287 ...
Page 293: ...Abdomen 293 ...
Page 307: ...Spine 307 ...
Page 311: ...Spine 311 ...
Page 321: ...Pelvis 321 ...
Page 327: ...Pelvis 327 ...
Page 338: ...338 Upper Extremities Hints For image reconstruction of soft tissue use kernel B30s B31s ...
Page 339: ...Upper Extremities 339 ...
Page 342: ...342 Upper Extremities Hint For image reconstruction of soft tissue use kernel B30s B31s ...
Page 343: ...Upper Extremities 343 ...
Page 357: ...Lower Extremities 357 ...
Page 369: ...Vascular 369 ...
Page 375: ...Vascular 375 ...
Page 379: ...Vascular 379 ...
Page 383: ...Vascular 383 ...
Page 387: ...Vascular 387 ...
Page 393: ...Vascular 393 ...
Page 397: ...Vascular 397 ...
Page 401: ...Vascular 401 ...
Page 405: ...Vascular 405 ...
Page 409: ...Vascular 409 ...
Page 413: ...Vascular 413 ...
Page 421: ...Specials 421 ...
Page 423: ...Specials 423 ...
Page 435: ...Specials 435 ...
Page 455: ...Specials 455 CARE Vision and Biopsy layout for CAREView equal size display Biopsy two segment ...
Page 461: ...Specials 461 ...
Page 469: ...Radiation Therapy 469 ...
Page 483: ...Radiation Therapy 483 ...
Page 517: ...Children 517 ...
Page 521: ...Children 521 ...
Page 525: ...Children 525 ...
Page 529: ...Children 529 ...
Page 533: ...Children 533 ...
Page 537: ...Children 537 ...
Page 545: ...Children 545 ...
Page 553: ...Children 553 ...
Page 557: ...Children 557 ...
Page 561: ...Children 561 ...
Page 565: ...Children 565 ...
Page 573: ...Children 573 ...
Page 577: ...Children 577 ...
Page 581: ...Children 581 ...
Page 585: ...Children 585 ...