Explosion protection 7MF1567-....1-....
Intrinsic safety "i" (only with current output)
•
II 1/2 G Ex ia IIC T4 Ga/Gb
•
II 1/2 D Ex ia IIIC T125 °C Da/Db
EC type-examination certificate
SEV 10 ATEX 0146
Connection to certified intrinsically-safe resistive circuits with
maximum values
U
i
≤
30 VDC; I
i
≤
100 mA; P
i
≤
0.75 W
Effective internal inductance and capacitance for versions
with connections to EN 175301-803-A and M12
L
i
= 0 nH; C
i
= 0 nF
Explosion protection to CSA 7MF1567-....1-....-Z; E21
Certificate of Compliance
70006348
Identification IS; Ex n; Ex ic; A/Ex nA
CL I, DIV 1, GP ABCD T4 ... T6; CL II, DIV 1, GP EFG; CL
III; Ex ia IIC T4 ... T6: CL I, DIV 2, GP ABCD T4 ... T6; CL II,
DIV 2, GP FG; CL III
Permissible ambient temperature
Ta = T4: -40 ... +85 °C (-40 ... +185 °F)
9
Dimensional drawings
9.1
Dimensional drawings electrical connections
M12x1/Fixcon
M16x1.5 or 0.5-14 NPT
9.2
Dimensional drawings process connections
Figure 9-1
Max. 30 Nm