For installations in which both the Ci and Li of the intrinsically safe apparatus exceeds1% of
the Co and Lo parameters of the MASS 6000 (excluding the cable), then 50% ofCo and Lo
parameters are applicable and shall not be exceeded. The reduced capacitanceshall not be
greater than 1
μ
F for Groups IIB and IIA, and 600 nF for Group IIC.
x
If the sensor, to which MASS 6000 is connected, measures liquids with temperatureabove
50 °C (122 °F), the MASS 6000 must be thermally insulated or separately installed.
x
The 24 V supply voltage shall be from a safety-isolated transformer according to EN 60742
RU(17KHSRZHUVXSSO\FDEOHZLUHVL]HPP
2
/ 24-14 AWG.
x
All the wiring for increased safety shall be in the separate section of the terminal compart-
ment and shall be covered with the plastic cover. The cover is marked with "Increased
Safety".
x
The transmittershall always be connected to PE (earth ground) using the internalterminal or
the external PE (earth ground).
x
The equipment shall either be installed in an area where there is a low risk of mechanical
damage or, if this not possible, the window shall be protected with an additional guard that
provides it with mechanical protection adequate for the application.
x
The equipment must be protected against direct sunlight and other ultraviolet light sources.
Directives
,QVWDOODWLRQ
7UDQVPLWWHU0$66([GFRPSDFWRUUHPRWHPRXQWHG
0539 II 2 G
DEMKO 03 ATEX 135253X
Ex d e ib [ia Ga] IIC T4 Gb
ambient temperature
20 to +50
q
C/
4
q
F to 122
q
F
EN 60079-0:2012+A11:2013
EN 60079-1:2007
EN 60079-7:2007
EN 60079-11:2012
EN 60079-25:2010