Figure 2-3
19" transmitter
Remove the transmitter from the installation and connect it to the SITRANS F M MAGFLO
Verificator as shown. Plug the Verificator adaptor into the normal transmitter location.
Once verification is completed, reinstall the transmitter. All application specific settings are
restored after verification.
Note
The Verificator requires no external power supply during verification and it automatically
establishes the correct power supply for each installation.
Connecting
SITRANS F M Verificator
8
Operating Instructions, 07/2019, A5E45134218-AB