![Siemens SITRANS DS III series Compact Operating Instructions Download Page 39](http://html.mh-extra.com/html/siemens/sitrans-ds-iii-series/sitrans-ds-iii-series_compact-operating-instructions_384652039.webp)
SITRANS P, DS III series (7MF4.33.. 7MF4.34.. 7MF4.35..)
A5E03434631-02, 12/2012
37
Absolute pressure input, with flush-mounted diaphragm
HART
PROFIBUS PA or Foundation Fieldbus
Measured variable
Gauge pressure
Span
Maximum
operating
pressure
MAWP (PS)
Maximum test
pressure
Measuring
range
Maximum
operating
pressure
Maximum test
pressure
43 …
1300 mbar a
(17 ...
525 inH
2
O)
2.6 bar a
(37.7 psi a)
10 bar a
(145 psi a)
1.3 bar a
(18.9 psi a)
2.6 bar a
(37.7 psi a)
10 bar a
(145 psi a)
160 …
5000 mbar a
(2.32 …
72.5 psi a)
10 bar a
(145 psi a)
30 bar a
(435 psi a)
5 bar a
(72.5 psi a)
10 bar a
(145 psi a)
30 bar a
(435 psi a)
1 … 30 bar a
(14.5 …
435 psi a)
45 bar a
(653 psi a)
100 bar a
(1450 psi a)
30 bar a
(435 psi a)
45 bar a
(653 psi a)
100 bar a
(1450 psi a)
Span (continuously
adjustable) or measuring
range, max. operating
pressure and max. test
pressure
Depending on the process connection, the
span may differ from these values
Depending on the process connection, the
measuring range may differ from these values
DS III input with PMC connection
HART
PROFIBUS PA or Foundation Fieldbus
Measured variable
Gauge pressure
Span
Maximum
operating
pressure
MAWP (PS)
Maximum
test
pressure
Measuring
range
Maximum
operating
pressure
Maximum
test pressure
0.01 … 1 bar g
(0.15 …
14.5 psi g)
1)
4 bar g
(58 psi g)
6 bar g
(87 psi g)
1 bar g
(14.5 psi g)
1)
4 bar g
(58 psi g)
6 bar g
(87 psi g)
0.04 … 4 bar g
(0.58 …
58 psi g)
7 bar g
(102 psi g)
10 bar g
(145 psi g)
4 bar g
(58 psi g)
7 bar g
(102 psi g)
10 bar g
(145 psi g)
Span (continuously
adjustable) or measuring
range, max. operating
pressure and max. test
pressure
0.16 …
16 bar g (2.3
… 232 psi g)
21 bar g
(305 psi g)
32 bar g
(464 psi g)
16 bar g
(232 psi g)
21 bar g
(305 psi g)
32 bar g
(464 psi g)
1)
1 bar g (14.5 psi g) only in PMC Style Standard, not in Minibolt
Absolute pressure input (from the gauge pressure series)
HART
PROFIBUS PA or Foundation Fieldbus
Measured variable
Absolute pressure
Span
Maximum
operating
pressure
MAWP (PS)
Maximum
test
pressure
Measuring
range
Maximum
operating
pressure
Maximum
test pressure
8.3 …
250 mbar a (3
... 100 inH
2
O)
1.5 bar a
(21.8 psi a)
6 bar a
(87 psi a)
250 mbar a
(100 inH
2
O)
1.5 bar a
(21.8 psi a)
6 bar a
(87 psi a)
Span (continuously
adjustable) or measuring
range, max. operating
pressure (in accordance
with 97/23/EC Pressure
Equipment Directive) and
max. test pressure (in
accordance with
DIN 16086)
43 …
1300 mbar a
(17 ...
525 inH
2
O)
2.6 bar a
(37.7 psi a)
10 bar a
(145 psi a)
1.3 bar a
(18.9 psi a)
2.6 bar a
(37.7 psi a)
10 bar a
(145 psi a)
Summary of Contents for SITRANS DS III series
Page 2: ......
Page 3: ...1 English 3 Eesti 50 Latviešu 96 Lietuvių 142 Polski 190 Română 239 ...
Page 4: ......