MOBILETT Plus HP
SPR8-220.091.01
Page 2 of 6
Siemens-Elema AB
Service
Rev. 03
03.03
SPS-UD
Solna, Sweden
2 - 2
Replacing the Inverter
3.
Open the holding belts (1/Fig. 2) and open the battery boxes by loosen three
screws (2/Fig. 2).
4.
Replace the old battery boxes.
Fasten the old top covers and tighten the holding belts.
Do not loosen the jumpers (1/Fig. 3) between the batteries.
Very high energy is stored in the batteries.
Life-threatening electric shock hazard exists.
MOB00353
Fig. 2
2
2
2
1
WARNUNG
WARNING
MOB00110
Fig. 3