Recalibration
7.4 Recalibration
BW500 and BW500/L
Operating Instructions, 12/2016, A5E33482052-AD
77
7.4.5
Routine Span
Note
To obtain an accurate and successful calibration, ensure that the required criteria are met.
Refer to Calibration Criteria (Page 199).
Press the following key:
Span Calibration. Current Span
Setup test. Press ENTER to start
41285
For example: the current span count
Zero should be done prior to Span
Setup test. Press Enter to start.
Perform a zero calibration or press:
Press the following key:
Span Calibration in progress
Current Reading:
%
55.56 kg/m
The load reported while calibration is in progress.
Calibration complete. Deviation
Press ENTER to accept value
0.03
41440
For example: the deviation in % from previous span, if the new span count is accepted. If unaccepta-
ble, press the following key to restart:
A deviation that is high (Siemens recommends three or higher) should be investigated further to
identify the source).
Note
During this step the progress of the process will be shown on the display with a progress
indicator in %.
Summary of Contents for Milltronics BW500
Page 1: ...Operating Instructions Milltronics BW500 and BW500 L 12 2016 Edition Integrators ...
Page 12: ...Table of contents BW500 and BW500 L 10 Operating Instructions 12 2016 A5E33482052 AD ...
Page 14: ...Safety Notes BW500 and BW500 L 12 Operating Instructions 12 2016 A5E33482052 AD ...
Page 16: ...The Manual BW500 and BW500 L 14 Operating Instructions 12 2016 A5E33482052 AD ...
Page 48: ......
Page 154: ......
Page 218: ...Index BW500 and BW500 L 216 Operating Instructions 12 2016 A5E33482052 AD ...
Page 219: ......