Installation
3 - 3
Siemens-Elema AB
Register 5
SPB7-230.092.11
Page 3 of 8
MAMMOMAT 3000 Modular
Solna, Sweden
Rev. 01
04.98
SPS-UD
Service
4.
Cut the cable ties at the two rear mounting screws for the X-ray tube.
5.
Remove the four mounting screws (13) to loosen the X-ray tube (see fig 3.2).
6.
Lift the X-ray tube at the front to access the filter-assembly motor.
7.
Loosen the screws for the filter-assembly motor, but do not remove them. Adjust
the position of the motor so that the filter assembly can rotate freely without
jamming (see fig 3.3).
8.
Tighten the screws for the motor.
9.
Reposition the X-ray tube paying attention to the guiding pin.
10.
Tighten the four mounting screws of the X-ray tube.
11.
Provide new cable ties at the rear mounting screws of the X-ray tube.
12.
System ON.
13.
Check the operation of the filter-assembly motor by selecting different anode/filter
combinations on the control panel.
Do not disconnect any cables.
NOTICE
Fig 3.3 Adjusting the filter-assembly motor