Troubleshooting with DIFS16P
R630/R640; M440/M450
TD00-400.840.10.03.02
Siemens
02.08
CS SD SP/CO
Page 24 of 32
Medical Solutions
Run diagnosis level 2
5.2
By choosing the diagnosis in “Level2”, the computer will be tested within approx. 30 min-
utes.
Fig. 6:
SC4 - Test and level selection for a user-configured diagnosis
"Level 2" scans more deeply than “Level1” but is more time consuming.
DIFS then shows the result of the diagnosis with ECS-Code, the location of the error, and
additional details of the failure.
If a defect could not be found in “Level2”, you can also select a single test in the Diagno-
sis/Configure dialog
. This test takes a minimum of one hour and can be set
to run repeatedly.