11/25
Building Technologies Division
CC1N7451en
20.11.2017
Technical data (cont'd)
Voltage at the ionization probe
- Operation
- Test
AC 330 V ±10 %
AC 380 V ±10 %
Short-circuit current
max. 0.5 mA
Recommended range of measuring
instrument
0...50 µA
Perm. length of detector cable
- Normal cable, laid separately ²)
- Shielded cable
max. 80 m
max. 140 m (e.g. high-frequency cable;
shielding connected to terminal 22)
Required detector current in operation
min. 6 µA
Possible detector current in operation
max. 200 µA
Supply voltage
- Operation
- Test
AC 330 V ±10 %
AC 380 V ±10 %
Required detector current
min. 70 µA
Possible detector current
- Operation
- Test
max. 700 µA
max. 1000 µA ¹)
Perm. length of detector cable
- Normal cable, laid separately ²)
- Shielded cable
max. 100 m
max. 200 m (e.g. high-frequency cable;
shielding connected to terminal 22)
¹)
During the prepurge time with higher test voltage: Self-ignition and extraneous light test
²)
Multicore cable not permitted
Detector current measurement
Ionization probe
Flame detector QRA2 / QRA4 / QRA10
LFL1...
24
7451v01/0204
M
C
+
+
-
ION
7451v02/0204
A
M
C
23 22
LFL1...
-
+
-
+
QRA...
For detector currents, refer to «Technical data».
C
Electrolytic condenser 100...470 µF; DC 10...25 V
ION Ionization
probe
M
Microammeter Ri max. 5,000
Flame supervision with
ionization probe
Flame supervision with
flame detector QRA2 /
QRA4 / QRA10
Measuring circuit for
detector current
measurement
Legend