Medical Solutions
•
Attach the two limit switches so that the cam enables the switching mechanisms at the
switches
.
•
When set correctly, you can hear the click emitted by the limit switches.
•
Angulation 0°.
•
Test the functioning of the limit switches by moving the C-arm into the +95° und -95° or-
bital positions.
•
Check these settings at least twice.