Redefine measuring ranges
G
G
The full-scale values can be set as desired. The
only condition is that the start-of-scale value set
using function 11 is greater than the full-scale
value set here.
The unit displayed when the setting is made
applies as the unit to the value entered here
(switch over unit using function 94).
Factory setting: 30/300/3000/30000
** 12 Define full-scale values of measuring ranges
CL
Range
2
1
Full-scale value
The start-of-scale values can be set as desired.
The only condition is that the full-scale value set
using function 12 is greater than the start-of-sca-
le value set here.
The unit displayed when the setting is made
applies as the unit to the value entered here
(switch over unit using function 94).
Factory setting: 0
** 11 Define start-of-scale values of measuring
ranges
CL
Range
1
1
Start-of-scale value
The number of valid measuring ranges is se-
lected. Autoranging (function 20) only takes
place within the selected ranges.
Identifier:
1 = measuring range 1
2 = measuring ranges 1 and 2
3 = measuring ranges 1, 2 and 3
4 = measuring ranges 1, 2, 3 and 4
Factory setting: 4
CL
n
Identifier
5
1
** 51 Select number of valid measuring ranges
A-26