
Display and switch over measuring ranges, switch over display
The LED on the right of the display indicates the
concentration unit of the display.
Factory setting: 0
The start-of-scale and full-scale values of the
measuring range are output in succession in the
display.
There are four measuring ranges. The measu-
ring range which is switched on is indicated to
the right of the display. The number of active
measuring ranges can be defined using function
51.
The start-of-scale and full-scale values of the
measuring range can be freely selected.
Factory setting: 4
A
* 94 Switch over between mg/m
3
and ppm
A
ppm C
3
mg C/m
3
3
9
2
CL
4
1
0
ppm C
6
ppm C
1
*
1 Change measuring range
Measuring range No.
CL
1
2 Display measuring range
Display
Measuring range No.
CL
2
This function only applies to the recorder output
(see plug connector X1). The current measured
value is always output on the display even if it
exceeds the current measuring range.
The displayed value flashes when the measuring
range(s) is exceeded.
Factory setting: 1
** 20 Autoranging on/off
Manual switchover
Autoranging
0
2
1
CL
0
A-14
Summary of Contents for 7MB1420
Page 72: ...A 36 ...
Page 73: ...A 37 ...