12
8019969/V1-0/2016-06 | SICK
A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S | S710 E
Subject to change without notice
2
ENGLISH
2.3
Change between “mg/m
3
” and “ppm”
Note
: Component oxygen (O
2
) is always shown in the physical unit % by vol. and cannot be
changed.
Fig. 3: Operating and display elements
1
Press “ESC” as often as required until you reach the main menu.
2
Press “.” (decimal point) to extend the menu selection.
3
Press key “6” to go to the “settings” menu.
4
A warning is displayed: Read the warning and confirm with “Enter”.
5
Press key “2” to go to the “measurement” menu.
6
Press key “3” for “meas. value display”.
7
Press key “1” to go to the “meas. component” menu. Select the measuring component
for which the physical unit is to be changed. Confirm your selection with “ENTER” and
exit the menu with “ESC”.
8
Press key “4” to go to the submenu where you can switch between “mg/m³” and “ppm”.