50
8013741/16LH/V1-7/2020-04| SICK
O P E R A T I N G I N S T R U C T I O N S | MCS300P Ex
Subject to change without notice
5
MENUS
5.5.1.4
Intensity
Menu 3.1.4: Diagnosis/Check values/Intensity
The intensities (energies) of the measured components are shown in this menu.
The amplification levels serve as information for SICK Customer Service.
5.5.2
Diagnosis/temperatures
Menu 3.2: Diagnosis/Temperatures
5.5.3
Diagnosis/system info
Menu 3.3: Diagnosis/System info
5.5.4
Diagnosis/error messages and Diag button
Menu 3.4: Diagnosis/Error messages
and
<Diag
button>
3.2
The actual temperatures are shown in this menu.
Status: Measuring
Back
Temperatures
/Diag/Temp
185 °C
xxx °C
60 °C
1 Cell
2 …
3 Optic hous
3.3
Device numbers and software versions are shown in this menu.
Status: Measuring
Back
System info
/Diag/Info
<name>
<1234>
<1234>
<1234>
1 System
2 SN-G
3 SN-K
4 System prc
●
Name of the system
●
Serial number of device
●
Serial number of cell
●
System software version
●
Etc.
3.4
Status: Measuring
Back
Error messages
/Diag/Err
System
S033 Temperature T1
too high
1/3 25/10 08:25:04
Existing
messages are shown in this menu (Logbook
→
SOPAS ET).
●
Date of occurrence (dd/mm)
Time of occurrence (hh:mm:ss)
●
Error cause (error number and plain-text)
(list of messages,
see “Error messages and possible causes”, page 69
)
●
Message number / number of existing messages
●
Source (e.g., system, measured component, receiver, etc. )