O P E R A T I N G I N S T R U C T I O N S | MCS300P Ex
Subject to change without notice
MENUS
5
5.4.2
Adjustment/automatic
Menu 2.2: Adjustment/automatic
5.4.3
Adjustment/Parameter
Menu 2.3: Adjustment/Parameter
2.2
34
Initiate automatic adjustments within in this menu item
(parameter settings
→
System Documentation).
The sequences of these programs depend on the parameter
settings of the programs (in the Figure: Examples).
For typical programs, status signals are set automatically and
valves for the test medium switched automatically.
For devices with internal adjustment standard (see System
Documentation): A corresponding program appears.
1
To start adjustment: Select the adjustment program and press “
Start
”.
2
Operating state:
“Maintenance”.
3
The Measuring screen is displayed (with a down counter to the end of
the adjustment).
4
After the end of the adjustment, the MCS300P Ex switches back to the
operating mode “Measuring” (If “
Maintenance
” was previously set
manually: Back to “
Maintenance
”).
●
If the deviation exceeds a limit (parameter settings in SOPAS ET), the
MCS300P Ex switches to classification
“Maintenance request”
(
“Status and classification”, page 36
).
Status: Measuring
Back
Start
Programs
/Adj/autom
1 Zero adjust all
2 Adjust Comp.1
3 Adjustment Comp.2
4 Program 4
Adjustment Comp.1
Menu
123 ppm
123 mg/m3
123 ppm
1 Comp.1
2 Comp.2
3 Comp.3
●
Down counter to the end of adjustment [seconds].
2.3
see “Concentration”, page 46
see “Factors”, page 47
see “Start times”, page 48
see “Start times”, page 48
Status: Measuring
Back
Enter
Parameter
/Adj/Par
1 Concentration
2 Factors
3 Start time 1 - 8
4 Start time 9 - 16