E. Field of view diagram
F. LED definitions
First level, runtime operation
Green
O
red
O
Yellow
O
Blue
O
Ready
Operation
Stop
Live
Busy
result
Decision OK
Decision NOK
-
-
Light
Lighting on
-
-
-
Function
User-defined
User-defined
User-defined
User defined
Second level, manual configuration (not active)
Green
O
red
O
Yellow
O
Blue
O
test
-
-
-
-
teach
-
-
-
-
setup
-
-
-
-
Focus
-
-
-
-
G. technical data
Field of view width x height in mm² (sq inch), typical values
0
300 x 225
(11.81 x 8.86)
200 x 150
(7.87 x 5.91)
100 x 75
(3.94 x 2.95)
0
200
(7.80)
100
(3.94)
300
(11.81)
Working distance in mm (inch)
Min. resolution in mm (mil)
1D code
2D code
0.1
(3.9)
0.2
(7.9)
0.05
(2.0)
0.05
(2.0)
0.2
(7.9)
0.3
(11.8)
0.1
(3.9)
a
b
c
d
e
f
g
h
i
j
Field of view width x height in mm² (sq inch), typical values
0
400 x 300
(15.75x 11.81)
600 x 450
(23.62 x 17.72)
800 x 600
(31.50 x 23.62)
1,000 x 750
(39.37 x 29.53)
1,200 x 900
(47.24 x 35.43)
200 x 150
(7.87 x 5.91)
0
200
(7.80)
400
(15.75)
600
(23.62)
800
(31.50)
1,000
(39.37)
1,200
(47.24)
1,400
(55.12)
1,600
(62.99)
1,800
(70.87)
2,000
(78.74)
2,200
(86.61)
Working distance in mm (inch)
Min. resolution in mm (mil)
1D code
2D code
0.1
(3.9)
0.2
(7.9)
0.4
(15.7)
0.6
(23.6)
0.8
(31.4)
1.0
(39.4)
1.2
(47.2)
1.4
(55.1)
0.2
(7.9)
0.3
(11.8)
0.4
(15.7)
0.5
(19.7)
0.6
(23.6)
0.7
(27.6)
0.8
(31.4)
0.9
(35.4)
1.0
(39.4)
a
b
c
d
e
f
g
h
i
j
1
2
3
4
Long range up to 2.2 m working distance
Short range up to 0.3 m working distance
attribute
Value
Features
Task
Reading
Technology
2D, snapshot, image analysis
Sensor
CMOS matrix sensor, gray scale
Spectral range
Approx. 400 nm ... 900 nm
LED class
White light: Risk group 1 (low risk,
IEC62471: 2006)
Laser class
1 (IEC 60825-1 (2014)) complies
with 21 CFR 1040.10 except for the
tolerance according to “Laser Notice
No.50” from June 24.2007
Focus
Manually adjustable
Lens
Exchangeable S- or C-mount, de-
pending on optical kit
Reading distance
30 mm ... 2000 mm
Performance
Scanning frequency
Max 50 Hz
Sensor resolution
1600 x 1200 px
Bar code types
Interleaved 2 of 5, Codabar, Code
128, Code 32, Code 39, Code 93,
GS1 DataBar, GS1-128/EAN 128,
Pharma, UPC/GTIN/EAN
2D code types
Data Matrix ECC200, GS1 Data-
Matrix, MaxiCode, QR code
Stacked code types
PDF417, Postal
Code qualification
On the basis of ISO/IEC 16022,
ISO/IEC 15415, ISO/IEC 15416,
ISO/IEC 18004
interfaces
Data store and
retrieve
Image and file logging via MicroSD
memory card, internal RAM and
external FTP
Switching inputs
Max 6 internal 24V (4 configurable
as in/out), additional 2 inputs via
optional CMC600 in CDB650/
CDM420
Configurable inputs
Encoder input, external trigger
Switching outputs
Max 4 internal 24V (configurable as
in/out) additional 2 external outputs
via CMC600 in CDB650/CDM420
Configurable
outputs
Good read, external light trigger,
conditional output
Output current
≤ 100 mA per output
attribute
Value
Maximum encoder
frequency
1 kHz
Control of external
illumination
Via digital output (24V trigger) or via
external light connector
Serial
- Function
- Rate
RS-232, RS-422
- Host, AUX
- 300 Baud ... 115.2 kBaud, AUX:
57.6 kBaud (RS-232)
USB
USB 2.0
Ethernet
- Function
- Data transmission
rate
- Protocol
- Host, AUX, image transmission
- 10/100/1000 Mbit/s
- TCP/IP, FTP
CAN bus
- Function
- Rate
- Protocol
- Master/Slave
- 250 kbit/s ... 500 kbit/s
- CSN (SICK CAN Sensor Network)
Optical indicators
11 LEDs (5 x status display, 5 x bar
graph, 1 green/red feedback spot)
Acoustic indicators
Configurable beeper
Control elements
2 function buttons
Memory card
MicroSD memory card, optional
Mechanics/electronics
Electrical connec-
tion
1 x M12, 17-pin plug (serial, CAN,
I/Os, power supply), 1 x M8, 4-pin
socket (USB), 1 x M12, 8-pin socket
(Ethernet), 1 x M12, 4-pin socket
(external light control)
Power consumption
Typ. 10 W, ± 20 %
Enclosure rating
IP 67 (with hood and plugged
unused connectors)
Housing
Aluminium
Window material
PMMA or glass
Weight
430 g excluding optics
Operating voltage
24 V DC ± 20 %
ambient data
Shock resistance
EN60068-2-27:2009-5
Vibration resistance
EN60068-2-6:2008:02
Ambient operating
temperature
0 °C ... +50 °C
Ambient storage
temperature
-20 °C ... +70 °C
Interpreting the diagrams
Use the diagrams to determine:
•
The maximum working distance
for a selected code resolution
•
The dimensions of the available
field of view
Example
Given (in
red
):
2D code resolution
1
: 1.0 mm
Read (in
green
):
Field of view
2
: 800x600 mm
2
Working distance
3
: 1400 mm
Lens focal length
4
: 12.5 mm
(example, other lenses possible)
a: f = 6.0 mm
b: f = 8.0 mm
c: f = 9.6 mm
d: f = 12.0 mm
e: f = 12.5 mm
f: f = 16.0 mm
g: f = 17.5 mm
h: f = 25.0 mm
i: f = 35.0 mm
j: f = 50.0 mm
Lens focal length