Commissioning and parameterization
FLOWSIC200 · Operating Instructions · 8013271/1CJ9/V2-0/2022-01 · © SICK Engineering GmbH
93
4.3.4
Bus addressing using SOPAS ET
As an alternative to addressing by hardware (
), the address can also be
assigned in SOPAS ET (
). To do this, connect the measuring system to SOPAS
ET, select device file “FLOWSIC200 M“, “FLOWSIC200 H” or “FLOWSIC200 H-M“ and set
the measuring system to “Maintenance” (
).
The miniature switch must be set to address 0 (
Fig. 83
“Configuration/Device Parameters“ directory
NOTICE:
For bus systems, it must be ensured that the bus addressing of the Master
FLSE200 is set correctly.
The sender/receiver units must have different addresses. Identical addresses
for several units cause the communication with the MCU to abort!
The default value for the bus address is always 1. Before connecting further
sender/receiver units to the bus, the ones already connected must first be
assigned a higher address.