Technical Data
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
195
16.6.2
Fuel gas
16.7
Sample gas conditions
Table 19
Specifications for the fuel gas
Fuel gas type:
1
1
Refer to entry in test log (explanation
P. 16, §1.4.3). Subsequent switchover not possible.
– H
2
: 100 %
H
2
– H
2
/He: Mixture
of
40 % H
2
+ 60 % He
Purity:
2
2
The fuel gas should not contain any hydrocarbons. Traces of CO and CO
2
are permitted.
For measuring range end values
10 ppm:
5.3
For measuring range end values > 10 ppm:
5.0
Source: Compressed
gas
cylinder
Table 20
Operating conditions for the fuel gas
Supply pressure:
1
1
at the gas inlet of the analyzer unit
300 kPa ±20 kPa (3.0 bar ±0,2 bar)(43.5 psig ±2.9 psig)
Volume flow during operation:
2
2
During ignition of the FID flame: Approx. 9 l/h
– H
2
:
1.3 … 1.8 L/h
3
3
For the exact value refer to the supplied test log
– H
2
/He:
●
Notes on selection of the correct fuel gas
●
Notes on installation
Materials of the components carrying the sample gas
Stainless steel 1.4301, graphite, quartz, platinum, sapphire, PTFE, Kalrez
Properties of the sample gas
Permissible sample gas
temperature:
– Extractive versions:
1
1
Analyzer unit types “wall housing” and “19" slide-in enclosure”
0 … 200 °C (32 … 392 °F)
– Inline versions:
2
2
Analyzer unit types “inline” and “industrial housing”
0 … 350 °C (32 … 662 °F)
Permissible dewpoint of the sample gas:
At ambient temperature
Permissible internal sample gas pressure:
3
3
Relative to the ambient/atmospheric air pressure
Atmospheric pressure ±5 kPa (±50 mbar)
Sample gas volume flow:
– Minimum:
40 l/h (670 cm
3
/min)
– Maximum:
170 l/h (2,800 cm
3
/min)
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...