162
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Maintenance
14.4
Servicing the sample gas filter
14.4.1
Access to the sample gas filter
Fig. 25
Sample gas filter: Position in the analyzer unit
●
Guideline for the maintenance interval of the sample gas filter (for
conventional applications): 2 months (1500 hours of operation)
●
If the sample gas filter is dismantled for maintenance work (e.g. if the filter
element was removed for cleaning), new sealing rings must be used for
reassembly (spare parts
).
Analyzer unit
Access to the sample gas filter
Type “wall housing”
Remove the cover.
Type “19" slide-in enclosure”
Open the front panel.
Type “inline”
Remove the outer housing.
Type “industrial housing”
Open the door.