OPERATING INSTRUCTIONS | DOSIC
®
8020955 /11HD/ 2018-12-12 | SICK AG
Subject to change without notice
32
7
COMMISSIONING
3. The password protection is activated using Activate.
NOTE
• The user is logged out again after 5 minutes of inactivity.
•
When the display is locked, only the configured measured value display can be seen.
7.5.11
Output options for Disp A and Disp B
Outside of the menu, the display alternates between different values, which can be
pre-configured via Disp A and Disp B.
•
Flow (Flow)
•
Fl+Unt (Flow and measuring unit)
•
Volume (Volume)
•
Temper (Temperature)
•
QaCurr (Current value)
•
QbCurr (Current value)
•
QxStat (Status of the output)
•
c (m/s) (Speed of sound)
•
v (m/s) (Flow velocity)
7.5.12
Selecting display unit
This setting enables you to configure the units used for flow, volume, and temperature.
Configuration
1.
Select the Config menu using the arrow pushbuttons and confirm using Set.
2.
Select the Units sub-menu using the arrow pushbuttons and confirm using Set.
Set values: Flow, Volume, Temper.
3.
Select the value and confirm using Set.
Depending on the parameter, the following units are available:
Parameter
Unit
Flow
• L/min
• L/h
• m
3
/h
• G/min
• Gal/h
Volume
• L
• m
3
• Gal
Temper
•
°C
•
°F