26
7 Storage
•
Always store the FC-M1 flow calibrator with the rubber feet oriented downward.
•
Store the unit within a temperature range of 0 to 40
°
C and humidity range of 10 to
90 % RH.
•
If the unit will not be used for an extended period, remove the batteries.
If automatic calibration fails, the screen shown to
the right is displayed.
If this screen appears, check the following and try
the operation again.
•
The dedicated communication cable or USB
cable is connected.
•
Power to the mini pump was not disrupted
during automatic calibration.
Automatic calibration is canceled even if power is
restored after a power interruption.
A USB error will occur if the dedicated
communication cable or USB cable is
disconnected during automatic calibration.
If the power supply to the mini pump is
interrupted during automatic calibration, start
over from step (1) in 6-5-1 Preparation for
Automatic Calibration.
For the MP-W5P model, if no pump operations
are performed for 30 seconds or more during
automatic calibration, the pump enters a sleep
mode that prevents communication. If that
occurs, press any key except the [START/STOP]
key on the MP-W5P unit to switch from the sleep
mode to the normal mode and enable
communication again.
FC-M1 screen after power
supply was interrupted to
the mini pump during
automatic calibration.