5-5 Temperature Display Mode Setting
The temperature used to convert flowrates can also be set manually (default setting is
[AUTO]).
→
Automatic setting
(based on temperature sensor values)
°
C
→
Manual setting
(initially displays temperature sensor value)
Press the [ENTER] key, which causes the display to flash, then use the [UP] or [DOWN]
key to select auto or manual mode. In the manual mode, press the [ENTER] key again
and then the [UP] or [DOWN] key to change the temperature value. Press the [ENTER]
key to confirm the setting. (To return to the previous screen without changing the
setting, press the [MODE/ESC] key.)
It is recommended that this setting be used only if there is a temperature
sensor problem. The setting reverts to [AUTO] when the power is switched OFF.
(If the indicated value is clearly not normal, request repairs.)
Automatic setting
Manual setting
(initially displays temperature sensor value)
(Confirm manual setting)
(numeric setting)
(Confirm)