- 38 -
Main Specifications
Item Code
090130-31
Model
HV-500R
Standard Suction Flow Rate
500 L/min. (with glass fiber filter
φ
110mm)
Configurable Flow Rate Range
100 to 800 L/min
Accuracy of Constant Flow Rate
Within
±
5 % of the set flow rate
Suction Pressure *1
-160 hPa (500 L/min)
Flow Rate Detection
Differential pressure detection method
Suction Pump
Brushless blower
Filter
φ
110 mm circular filter
Screen
Touch panel LCD (backlit)
Power Outage Countermeasures
Continues with the prior operating conditions after power recovery
Operating Temperature Range
0 to 40
°
C with no condensation
Power Supply
220VAC (180VAC to 264VAC), 50/60 Hz
Current Protection Function
15 A Circuit protector
Power Consumption *2
500 L/min: 1.4 A (with glass fiber filter
φ
110mm)
Dimension
425 (W)
×
200 (D)
×
270 (H) mm
Weight
Approx. 8 kg
Accessories
Filter clip for
φ
110mm (Item Code 080130-0871),
Glass fiber filter
φ
110mm GB-100R-110A 10pcs,
Touch panel protection sheet, Power cable
*1: Operation under 1 atmospheric pressure assumed. Performance may be inferior under low atmospheric
conditions at high altitudes.
*2: This is the initial value (at time of shipment, at AC220V). Power consumption may exceed double the
specified value if a collector with a large load exceeding the specifications is used.
Spare Parts
Item Code
Name of Part
Details
080130-0871
Filter clip (standard accessories)
For
φ
110mm
Consumables
Item Code
Name of Part
Details
080130-98110
Glass fiber filter, PTFE binding
φ
110mm, 50pcs, TF98
080130-031
Glass fiber filter
φ
110mm, 100pcs, GB-100R-110A
080130-034
Quartz fiber filter
φ
110mm, 100pcs, QR-100