- 31 -
*
The factory default span value is 1.000, and the zero value is
000.0.
Calibration Mode
In the menu screen, press [CAL] to switch to calibration mode. This section only describes the
operating methods in the calibration mode. When performing an actual calibration, see the section
on calibration methods. (See page 33.)
[SPAN] key
Pressing this key will proceed to the respective span value entry screens.
[ZERO] key
Pressing this key will proceed to the respective zero value entry screens.
Leak check
Pressing this key will proceed to the leak check screen. (See page 32.)
Blower operating time
Displays the operating time for the blower mounted in this product.
[MENU] key
Pressing this key will return to the menu screen.
[ESC] key
Pressing this key will return to the main screen.
The respective span value/zero value entry
screens will be as in the figure at right. Press a
numerical region to proceed to the numerical value
entry screen.
When numerical values have been entered, the
date of entry is recorded, and the system returns to
this screen.
Press [ESC] to return to the menu screen.
In the numerical value entry screen, press the
numerical keys to enter the desired numerical
values.
Press
to confirm the numerical values as the
setting values and return to the respective span
value/zero value entry screens.
Press [ESC] to cancel entry and return to the
respective span value/zero value entry screens.
Press [CLR] to set the numerical value to "0"
Press [BS] to delete the last figure entered.
[DEL] is not available here.
■SET FLOW SPAN
ESC
1 . 0 0 0
0 9 / 0 9 / 1 5 1 0 : 0 0 : 0 0
■SET FLOW SPAN
ES
C
0
5
9
8
4
3
7
2
B
S
CL
R
1 . 0 0 0
6
1
DE
L
.
Blower
operating time
[MENU] key
[ESC] key
Leak check
[SPAN] key
FLOW
PRES.
BLOWER 1000 hr 00 min
MENU ESC
LEAK CHECK
SPAN
SPAN
SPAN
ZERO
ZERO
ZERO
TEMP.
Flow rate
Atmospheric pressure
Atmospheric
temperature
[ZERO] key