background image

ADM-870 05/23/08

36

12.0  SPECIAL BALANCING PROCEDURES

12.1  PROPORTIONAL BALANCING

Backpressure compensated readings should be taken during the preliminary survey of the entire system with

all dampers fully open, and also during the final readings after balancing is complete.  Nonbackpressure

compensated readings require less time and are usually adequate for the proportional balancing of outlets (see

Section 5.4 SPEED-READ).

12.2  LARGE RETURN AIR GRILLES

Fan systems such as package air conditioning units or fan coil units may have only one main return air inlet.

This is common in residential air conditioning systems ranging from two tons to 7 ½ tons of cooling capacity.

The rated air delivery range is 800 to 3000 cfm.  The FlowHood may be used to measure the return air flow if

special attention is paid to avoiding excessive restriction of the air flow.  The flow restriction is kept to a

minimum by taking readings in two or more segments.  For instance, if the return air grille is 30" x 24" and rated

flow is 2400 cfm, read each half of the grille (15" x 24") using backpressure compensation.  The sum of the

backpressure compensated readings taken on each half is the total air flow through the return air grille.  Keep

in mind that this flow measurement represents only the flow at the return air grille.  The total air flow at the fan

includes return air duct leakage and possible outside air intake.

The reading for each portion of the grille would be even less restrictive and thereby more accurate if the 1'x 4'

top were used on the FlowHood.  The air flow could then be measured in three segments along the 30"

dimension, with the excess length of the 48" top overlapping onto the ceiling.  The 30" dimension of the grille

could be broken into 9", 12" and 9" segments using the 12" width of the 1'x 4' FlowHood top.  This procedure

is also useable on large supply or discharge grilles and diffusers.

12.3  KITCHEN EXHAUST HOODS

Exhaust fan delivery of room temperature air in kitchens is reduced substantially when the cooking surfaces,

ovens, broilers and fryers are heated to normal working temperatures. This is caused by the significant

reduction in air density which occurs during the change from cool set-up to heated working conditions.  

A fan which was set to deliver 5000 cfm at 75

/

 F intake air will actually deliver only about 4610 cfm of the 75

/

F room air when the air is heated to 140

/

 F in the exhaust hood.  It is very important that a kitchen system

balance be confirmed by measurements for air flow, room pressure, and actual makeup air, while the cooking

devices are at their normal working temperatures.

12.3.1  RANGE EXHAUST FILTERS AND GREASE EXTRACTORS 

IMPORTANT:  See Section 6.1 VELOCITY CORRECTION FACTORS

The VelGrid may be used to measure the range exhaust washable filter and grease extractor face velocities

as described below.    NOTE: The VelGrid is 

not

 designed to be used without the standoffs. 

Begin by determining the gross face area of each filter or grease extractor to be tested.  In this case, the

nominal filter size marked on the filter should be used to calculate the full gross face area.  For example, a 20"

x 25" filter is 500 inches square overall (or 3.47 feet square overall).  Measure the face velocity of each filter

or extractor with the VelGrid positioned at the center of the filter face only.  The 1.5" standoffs must be pressed

against the face of the filter to hold the VelGrid 1.5" from the filter face.  Determine the Kv factor for this type

of filter as described in Section 6.1 VELOCITY CORRECTION FACTORS.  Multiply the center face velocity

by the calculated Kv factor to obtain the corrected face velocity for that filter or extractor.  The air flow in cfm

or l/s is calculated by multiplying the gross face area of the filter or extractor by the 

corrected

 face velocity.

Most commercial grade exhaust hoods leak to some degree between the filters and along the filter support

tracks.  The sum of the filter air flows obtained in the manner described above is generally sufficiently accurate

for properly constructed kitchen exhaust hoods.  Hoods which have been installed with excessive clearances

and leakage will require correction for leakage. 

12.4  CONSTANT VOLUME CONTROLLERS

The backpressure compensation process requires special attention when used to measure air flow at

individually controlled constant volume air terminals.  The nonbackpressure compensated and the

backpressure compensated air flow readings should be essentially equal for fast acting constant volume air

terminals.  The controller responds to the additional backpressure and maintains the preset air flow.  If the

controller is slow acting, the FlowHood must be held in place long enough for the controller to reattain the set

point.

Master and slave, volume controlled outlets often do not interact favorably with the FlowHood backpressure

compensation function.  Air delivery measurement for this type system should be performed using a

nonbackpressure compensated measurement.

Summary of Contents for Airdata ADM-870

Page 1: ...nts Inc 7855 East Redfield Road Scottsdale Arizona 85260 3430 Phone 480 991 6744 Fax 480 443 1267 www shortridge com S i AIRDATA MULTIMETER ADM 870 ELECTRONIC MICROMANOMETER AIR FLOW VELOCITY PRESSURE...

Page 2: ...IL 18 6 3 3 EXHAUST HOODS AIRFOIL PROBE 19 6 3 4 LAMINAR FLOW WORKSTATIONS AIRFOIL PROBE 19 6 4 VELGRID AIR VELOCITY 19 6 4 1 CHEMICAL EXHAUST HOODS VELGRID 20 6 4 2 LAMINAR FLOW WORKSTATION VELGRID 2...

Page 3: ...L PRESSURE FUNCTION 40 14 3 ABSOLUTE PRESSURE FUNCTION 40 14 4 AIR FLOW ACCURACY 40 14 5 DUCT TRAVERSE COMPARISONS INCLINED MANOMETER MICROMANOMETER 40 14 6 DUCT TRAVERSE USING THE AIRDATA MULTIMETER...

Page 4: ...6 4 VELGRID ASSEMBLY 21 7 1 STATIC PRESSURE PROBE 23 8 1 ADT442 TEMPROBE 25 8 2 AIRDATA MULTITEMP 26 10 1 FRAME STORAGE 30 10 2 FLOWHOOD IN CASE 30 10 3 FLOWHOOD ASSEMBLY 31 10 4 2X2 FRAME ASSEMBLY 32...

Page 5: ...emperatures can be measured from 67 F to 250 F either individually or in conjunction with an air flow or velocity measurement The Model ADM 870 AirData Multimeter performs these additional functions w...

Page 6: ...readings AIR DENSITY CORRECTION The air density correction range is 14 40 in Hg and 32 F to 158 F for correction of air flow and velocity measurements The readings represent either local density air...

Page 7: ...e accurate measurements Bleedthrough typ 0 0004 cu in in wc measurement Quiescent bleedthrough max 0 0005 cu in in wc minute TUBING The maximum recommended length of pneumatic tubing for the measureme...

Page 8: ...emory mode then alternate action for average and for total This key was labeled MEM prior to May 1993 ON Turns the meter on RCL Recalls stored readings in reverse order RCL Recalls stored readings in...

Page 9: ...e meter is dropped and may cause the keypad to become nonfunctional until the meter is reset If the meter continues to fall into lockout it may have been damaged and should be returned for repair PNEU...

Page 10: ...EAD Read Hold for Speed or Halt RCL Recall in Entry Order RCL Recall in Reverse Order OFF Power Off ON Power On Positive Pressure Port Negative Pressure Port Flaps Jack for FlowHood Flaps Plug Threade...

Page 11: ...hat the temperature readings will be displayed in degrees Celsius C LS READ This display indicates that the meter has been placed in the air flow function and will read in litres per second L s S READ...

Page 12: ...mode is selected and also upon each subsequent operation of the READ key AUTOSTOR This message indicates that the meter has been placed in the automatic memory mode which integrates the automatic rea...

Page 13: ...de is selected or when the meter is first turned on if the flaps plug is connected HALT This message will be displayed when an automatic reading sequence has been halted manually by holding down the R...

Page 14: ...tions do not apply to SPEED readings STD 70 F This message will be flashed during air flow or velocity measurements performed without the TemProbe The resulting flow or velocity value will be calculat...

Page 15: ...ADM 870 05 23 08 11 VELGRID This signal will be flashed when the VelGrid mode is selected and also upon each subsequent operation of the READ key...

Page 16: ...ited by pressing the SHIFT key followed by the CLEAR key 5 2 MEMORY AVERAGE TOTAL FUNCTION The memory average total function may be used with any of the measurement modes and allows the storage of up...

Page 17: ...ay be combined to gain the increased resolution afforded by the increased time base of the measurement interval This function permits the operator to store up to 100 fully automatic repeated readings...

Page 18: ...he SPEED READ function may be terminated while HALT is displayed by pressing the SHIFT key followed by the CLEAR key or by turning the meter off The memory function may be activated while HALT is bein...

Page 19: ...ating air flow from velocity measurements at most air movement devices such as Clean Room HEPA filters chemical exhaust hoods safety cabinets laminar flow work stations coil and filter face velocities...

Page 20: ...90 degree angle located near the base The measurement range of the AirData Multimeter with the pitot tube is 25 to 29 000 fpm calibration accuracy is certified to 8 000 fpm The stainless steel pitot t...

Page 21: ...city pressure rather than velocity A pitot tube used with the AirData Multimeter reads out directly in velocity when used in the PITOTUBE mode and reads velocity pressure when used in the DIF PRES mod...

Page 22: ...tally and up against the bottom edge of the sash door Tape markers on the AirFoil probe and along the sash door edge at 6 centers will aid in accurate positioning of the AirFoil probe The individual r...

Page 23: ...nt is usually adequate 6 4 VELGRID AIR VELOCITY IMPORTANT See Section 6 1 VELOCITY CORRECTION FACTORS The VelGrid accessory is designed especially for use in the measurement of general face velocity c...

Page 24: ...panel face The velocity average obtained in this manner can be used to calculate the volumetric air flow rate as described in Section 6 4 3 AIR FLOW CALCULATION 6 4 3 AIR FLOW CALCULATION FROM VELGRID...

Page 25: ...ADM 870 05 23 08 21 6 4 VELGRID ASSEMBLY...

Page 26: ...ity conditions other than standard must be corrected for local air density conditions if these results are to represent true velocity Air velocity readings taken with the ADM 870 in the standard densi...

Page 27: ...RESSURE PROBES Static pressures are measured with static pressure probes These probes are brass colored and have a single tubing connection in the magnetic base Connect the tubing from the static pres...

Page 28: ...the atmosphere and the meter is in the absolute pressure mode The measurement range is 10 40 in Hg Specific absolute pressure source measurements may be obtained by connecting the pressure source dir...

Page 29: ...emp combines with the memory function of the AirData Multimeter to store up to 100 temperature readings along with the sequence tag for each reading Each temperature reading may be entered into memory...

Page 30: ...lugs or direct immersion Some applications include humidity control systems direct expansion A C systems outside air adjustment hot water heating chilled water condenser water and many other circulati...

Page 31: ...n the display immediately prior to the display of the air flow reading The flow is calculated using an assumed standard temperature of 70 F 9 2 BACKPRESSURE COMPENSATION The air flow delivery of a sup...

Page 32: ...nel when the two pieces are slid together Side extension channels are joined together by a thumbnut and splice angle arrangement The individual frame pieces are stamped with a frame number as shown ci...

Page 33: ...added to both the top and bottom of the support dowels The dowel end pins are to be inserted into the frame corner brackets as shown in Figure 10 8 Each frame piece has been labeled to indicate posit...

Page 34: ...ADM 870 05 23 08 30 FIGURE 10 2 FLOWHOOD IN CASE FIGURE 10 1 FRAME STORAGE...

Page 35: ...ADM 870 05 23 08 31 FIGURE 10 3 FLOWHOOD ASSEMBLY...

Page 36: ...ADM 870 05 23 08 32 FIGURE 10 4 2X2 FRAME ASSEMBLY FIGURE 10 5 1X4 FRAME ASSEMBLY FIGURE 10 6 2X4 FRAME ASSEMBLY...

Page 37: ...ADM 870 05 23 08 33 FIGURE 10 7 1X5 FRAME ASSEMBLY FIGURE 10 8 3X3 FRAME ASSEMBLY...

Page 38: ...her at any convenient point on the outside of the base Set the flaps control knob to the open position Press the READ key or the thumbswitch on the FlowHood handle and the reading will be displayed di...

Page 39: ...with the flaps open as described above Close the flaps when the meter displays CALC When the first reading is displayed press the READ key to take the second backpressure compensated reading If necess...

Page 40: ...in the exhaust hood It is very important that a kitchen system balance be confirmed by measurements for air flow room pressure and actual makeup air while the cooking devices are at their normal work...

Page 41: ...use an extreme concentration of air velocity on portions of the flow sensing grid The FlowHood readings will be inaccurate under such conditions Reading accuracy is improved by adjusting the register...

Page 42: ...the density effects of relative humidity using the formula shown below This calculation requires the use of the vapor pressure which must be looked up on a Dew Point Vapor Pressure conversion chart T...

Page 43: ...n with the ADM 870 AirData Multimeter may be compared with hot wire anemometer readings if the hot wire readings are corrected for local air density conditions to obtain true air velocity results The...

Page 44: ...ure source must be self replenishing such as a pressure drop across a needle valve orifice or orifice plate 14 3 ABSOLUTE PRESSURE FUNCTION Confirm the accuracy of the absolute pressure correction by...

Page 45: ...obe is not used for either the flow or velocity measurement it must not be used for the other type of measurement In this case both sets of measurements will be calculated using standard 70 F conditio...

Page 46: ...anything inside the meter will void the warranty and may cause serious damage to the sensitive electronic components The AirData Multimeter should be returned to the factory at least every two years...

Page 47: ...tional cushioning to fill the carton Do not use spray foam Spray foam can damage the meter and it is possible to lose a meter in a chunk of spray foam Ship directly to Shortridge Instruments Inc 7855...

Page 48: ...gular charges will apply This warranty shall be effective only if use and maintenance is in accordance with our instructions and written notice of a defect is given to the Seller within such period TH...

Page 49: ...for the density effect of variations in temperature Record the AirData Multimeter temperature and absolute barometric pressure reading at the beginning and end of each probe test range These readings...

Page 50: ...et up with different pressures of precisely the desired test differential pressure These testers must be carefully adjusted to an accurate equal base pressure of approximately 3 psi before adding weig...

Page 51: ...the upper half positive 6 cells of the battery set by placing your DC voltmeter common ground probe at the left center battery clip and the other probe at the upper right corner battery clip The volt...

Page 52: ...d above The temperature during storage should be less than 95 F The batteries will function best after storage if reconditioned by performing two full cycles of 24 hour charge and full discharge follo...

Page 53: ...e IM 870 1 Instruction manual ADM 870 8400 FLOWHOOD 844 1 FlowHood base assembly with flaps 845 1 Flow sensing grid 826 1 Top support assembly 842 1 Pushbutton handle assembly 827 4 Support dowels 838...

Page 54: ...set complete 3 x 3 TOP 91 4 x 91 4 cm 141 4 Side channel type 1 336 4 Side channel type 6 3x3F 1 Complete frame assembly all of above 3x3S 1 Cloth skirt 36 x 36 827E 8 Support dowel extenders 3x3T 1...

Page 55: ...nnn 7 chemical exhaust hoods 19 clean room filter outlets 19 CLEAR 4 9 12 clear memory 13 constant volume controllers 36 correction factors 18 38 default air density correction 18 DENS 4 density corre...

Page 56: ...all 6 RECHARGE 10 relative humidity 38 relative humidity correction 38 replacement parts list 49 50 reset switch 6 response time 2 safety cabinets 19 safety notice 34 sea level conditions 22 Sheet Met...

Reviews: