Shortridge Instruments ADM-860C Operating Instructions Manual Download Page 57

ADM-860C 04/14/08

52

14.0  METER ACCURACY FIELD TESTING

14.1  METER ZERO FUNCTION

Disconnect all tubing from the positive (+) and negative (-) ports of the meter.  Perform several readings with the
meter set for the flow or velocity mode, with 

no air

 passing across the meter ports.  The meter should display

zero readings, but may occasionally show a low reading such as 25 cfm or fpm.  An 

occasional

 positive or

negative reading is random electronic "noise" and may be disregarded.

If the zero test readings tend to be either positive or negative and greater than 25 cfm, the zero bias indicated
will affect the accuracy of low air flow or velocity measurements.  Check for any obstruction of the meter
connections and clean if needed.  

Be sure that there is no moisture or condensation in the meter or meter connections.  If the meter has been
exposed to water or high humidity, followed by rapid cooling, water may have condensed inside the meter.  In
this event, the meter should be placed in a warm dry atmosphere (between 80

/

 F and 130

o

 F) for 24 hours.

Afterward, the batteries will need to be recharged for 10 hours at a temperature less than 113

o

 F.  If a zero bias

or other problem persists, call the factory for advice.

14.2  DIFFERENTIAL PRESSURE FUNCTION

Check the differential pressure function zero accuracy by taking a reading with both meter ports open to the
atmosphere.  Be sure the ports are 

not

 exposed to a draft.  The reading should be 0.0000 in wc ± 0.0001 in wc.

Check the differential pressure reading accuracy by measuring a constant pressure source at approximately the
following pressures:  0.1250 in wc; 2.000 in wc; 30.00 in wc.  Confirm the pressure readings by comparison with
an appropriately accurate inclined manometer at the 0.1250 and 2.000 in wc differential.  The 30.00 in wc
differential may be checked using a U-tube and water.

Since the AirData Multimeter passes a very small sample of the air through the meter during each measurement
(see SPECIFICATIONS), it should be connected to the pressure source independently of any other meter,
manometer, or U-tube.  To avoid pulsation or "cross talk" through the pressure source tubing, use shutoff valves
to isolate the AirData Multimeter and the reference pressure gage.  The pressure source must be self-
replenishing, such as a pressure drop across a needle valve, orifice, or orifice plate.

14.3  ABSOLUTE PRESSURE FUNCTION

You may confirm the accuracy of the absolute pressure correction by taking a reading with the meter ports open
to the atmosphere and comparing the reading with the actual barometric pressure.  The reading should be within
± 2% (approximately 0.5 in Hg) of the actual barometric pressure to maintain the specified accuracy for velocity
and air flow density correction.  

CAUTION:  Testing of absolute pressures greater or less than local barometric pressure must be performed with
the reference pressure applied to the positive (+) and the negative (-) ports at the same time.  This precaution
avoids excessive pressure input to the differential pressure transducer.

14.4  AIR FLOW ACCURACY

Air flow accuracy is confirmed by comparing a very careful pitot tube traverse with the results obtained using the
FlowHood.  An accurate inclined manometer, micromanometer, or AirData Multimeter may be used for the duct
traverse.  An inclined manometer used for the duct traverse should have minor scale divisions of 0.005 in wc or
less, and should have direct velocity markings down to at least 400 fpm.  The accuracy of the comparison tests
will depend on both the accuracy of the velocity traverse, and the accuracy of the air flow calculations.

A multipoint pitot tube traverse is performed on a supply duct which serves a single supply diffuser.  The pitot tube
should be a type approved by the NPL (National Physics Laboratory, U.K.) or AMCA (Air Moving and
Conditioning Association, U.S.).  The duct velocity should be at least 800 fpm; the duct should be properly sealed
and taped; and the connection to the diffuser should be airtight.  The duct traverse location should have straight
duct for six to eight duct diameters upstream, and three to four diameters downstream.  For example, a one foot
diameter duct requires eight feet of straight duct upstream and four feet downstream.

Summary of Contents for ADM-860C

Page 1: ...st Redfield Road Scottsdale Arizona 85260 3430 Phone 480 991 6744 Fax 480 443 1267 www shortridge com S i AIRDATA MULTIMETER ADM 860C ELECTRONIC MICROMANOMETER WITH RS232 COMMUNICATIONS PORT AIR FLOW...

Page 2: ...5 INDIVIDUAL READINGS CONTROLLED FROM COMPUTER KEYBOARD 23 5 5 6 DOWNLOAD READINGS FROM MEMORY USING COMPUTER KEYBOARD 23 5 5 7 CONTROL BUTTONS CURSOR CONTROL AND FILE PATHS 23 5 5 8 AUTOMATIC REPEAT...

Page 3: ...2 7 14 x14 x14 SHORT TOP SET 49 12 8 SYSTEM PROBLEMS 49 13 0 CORRECTION FACTORS 50 13 1 BAROMETRIC PRESSURE DENSITY CORRECTION LOCAL DENSITY 50 13 2 BAROMETRIC PRESSURE DENSITY CORRECTION STANDARD DEN...

Page 4: ...EDURE 59 PROLONGING BATTERY LIFE 60 BATTERY RECYCLING 60 REPLACEMENT PARTS LIST 61 INDEX 63 Copyright Shortridge Instruments Inc 2008 All rights reserved This information may not be reproduced or dupl...

Page 5: ...6 3 VELGRID ASSEMBLY 33 7 1 STATIC PRESSURE PROBE 34 8 1 ADT442 TEMPROBE 36 8 2 AIRDATA MULTITEMP 37 10 1 FRAME STORAGE 42 10 2 FLOWHOOD IN CASE 42 10 3 FLOWHOOD ASSEMBLY 43 10 4 2X2 FRAME ASSEMBLY 4...

Page 6: ...er features sequential storage and recall of up to 100 individually obtained measurements with sequence tags for each value Recall of the sum and average along with the minimum and maximum readings fo...

Page 7: ...o 2000 cfm nonbackpressure compensated readings AIR DENSITY CORRECTION The air density correction range is 14 40 in Hg and 32 F to 158 F for correction of air flow and velocity measurements The readin...

Page 8: ...24 Volts AC to the meter Batteries may be left on charge for an unlimited time without harm The temperature of the instrument during charge should be kept between 40 F and 113 F 5 C to 45 C The meter...

Page 9: ...m readings ON Turns the meter on Turns light on and off after meter has been turned on RCL Recalls stored readings in reverse order RCL Recalls stored readings in entry order READ Initiate measurement...

Page 10: ...switch down or press the switch twice in close sequence PNEUMATIC PRESSURE INLETS Two pneumatic pressure inlets positive and negative are centered on the back of the meter at the top edge and may be c...

Page 11: ...all in Entry Order RCL Recall in Reverse Order OFF Power Off after Shift ON Power On Turn Light On Off Positive Pressure Port Negative Pressure Port Flaps Jack for FlowHood Flaps Plug Threaded Insert...

Page 12: ...cates that the absolute pressure readings will be displayed in bars 1 bar 100 kPa with reference to a vacuum C READ Display indicates that the temperature readings will be displayed in degrees Celsius...

Page 13: ...r backpressure effects Pa nnn nn Indicates that the result represents a differential pressure measurement Pa MS nnn nn Indicates that the displayed result represents a velocity measurement m s 4 3 FUN...

Page 14: ...emory The display shows the reading and its place in the reading sequence during a STORE process Accessed using forward RCL keys nnu nnnn Displays the number of nonbackpressure compensated readings in...

Page 15: ...dicates the units for the reading xx n nnnn Displays the maximum greatest value reading stored in memory during a STORE sequence Displayed when the STORE key is pressed four times after HALT during th...

Page 16: ...ains in the batteries BATT 2 3 This message may be displayed following SHIFT SHIFT and indicates that about two thirds of the useful charge remains in the batteries BATT FULL This message is displayed...

Page 17: ...is message indicates that the display back light has been turned off Battery time is extended when the back light is turned off LIGHT ON This message indicates that the display back light has been tur...

Page 18: ...e measurement This message is also displayed if the TemProbe or extension cord has been damaged so as to create an open circuit NOT ALLOWD This message is displayed if the user tries to select STORE m...

Page 19: ...message is displayed during differential pressure measurements It is also displayed during the manual reading which is taken when a set of TREND readings is halted RECHARGE This message signals that...

Page 20: ...O COLD and shut down However if there are readings in memory the meter will continue to display TOO HOT or TOO COLD and will retain the readings in memory The meter must be cooled down or warmed up as...

Page 21: ...ack light may be turned on or off by pressing the ON OFF key while the meter is turned on The display will read LIGHT ON or LIGHT OFF as appropriate The third character colon in each READ display is r...

Page 22: ...djusted approaches the set point TREND mode sequential readings are optimized for speed not accuracy When the air has stabilized near the required setting the meter can be switched to the MANUAL readi...

Page 23: ...rage capacity of 100 readings has been reached STORE FULL will be displayed Press the STORE key to alternately display the average and sum or one of the RCL keys for recall of the individual readings...

Page 24: ...gns with the red rectangle on the lower part of the jack Connect the null modem adapter to the DB9 connector on the serial cable and then plug the null modem adapter into the serial port on the printe...

Page 25: ...nd manipulation This functionality requires either HyperTerminal Private Edition or WinWedge software from TAL Technologies Inc as well as an RS232 serial cable The procedure using HyperTerminal is su...

Page 26: ...e Send Keystrokes To Port Settings Define Input Data Record Structure Pre Input Character Translations Pre Transfer Character Translations Serial Output Strings Hot Keys and Hot Key Actions The Mode S...

Page 27: ...mmand or a series of commands to the meter The meter recognizes the command and responds by changing modes taking a reading or downloading stored readings from memory The function keys are set up in t...

Page 28: ...guration file to be used Position the cursor in the cell on the spreadsheet where data entry should begin Select the appropriate mode on the meter using the predefined computer keyboard function key s...

Page 29: ...ng immediately after each previous reading is completed Automatic readings will continue until the user selects Quit Quit or Quit Suspend Wedge in the WinWedge window Readings may also be programmed t...

Page 30: ...D It is possible to use the printer download methods discussed in Sections 5 4 1 DOWNLOAD INDIVIDUAL READINGS TO A PRINTER and 5 4 2 DOWNLOAD READINGS IN MEMORY TO A PRINTER while connected to a compu...

Page 31: ...ons The use of an Ak factor is not appropriate however in the measurement of face velocities work zone velocities or in calculating air flow from velocity measurements at most air movement devices suc...

Page 32: ...nding five step procedure seems to leave little room for the art of Testing and Balancing This is not altogether true The measurement of the air velocity in Step 4 is affected by the position and orie...

Page 33: ...lts Pitot tubes are available in several different sizes and configurations to simplify different applications which may be encountered When a pitot tube is used in internally insulated ducts small pa...

Page 34: ...ement range of the AirFoil probe is 25 to 5 000 fpm The AirFoil probe is useful for free point air velocity measurements such as exhaust hood face velocities HEPA filters or laminar hood velocities A...

Page 35: ...1 VELOCITY CORRECTION FACTORS The exhaust filter face velocity may also be tested with the AirFoil probe to determine exhaust air flow The cabinet manufacturer s probe position schedule should be used...

Page 36: ...easurement range using the VelGrid is 25 to 2500 fpm Each reading represents 16 velocity points over a 14 x 14 area 1 36 sq ft The VelGrid unit is assembled by attaching the pushbutton handle to one o...

Page 37: ...re placed directly against the perforated supply panel face The velocity average obtained in this manner can be used to calculate the volumetric air flow rate as described in Section 6 5 3 AIR FLOW CA...

Page 38: ...ADM 860C 04 14 08 33 FIGURE 6 3 VELGRID ASSEMBLY...

Page 39: ...open to the ambient air pressure Place the end of the tubing in one area and place the meter in the other The meter will measure the pressure differential between the two areas 7 1 1 STATIC PRESSURE...

Page 40: ...tube into the airstream as discussed under Section 7 1 2 PITOT TUBE VELOCITY PRESSURES above The resulting differential pressure reading represents total pressure 7 2 ABSOLUTE PRESSURE The absolute p...

Page 41: ...exposed to temperatures beyond this range The maximum recommended safe exposure range is 100 F to 250 F and the probe accuracy may be effected by exposure beyond this range Do not expose the plastic b...

Page 42: ...e by using more than one MultiTemp switch unit connected in series piggybacking Reading and memory entry of 64 temperature points would take about two minutes after temperature probes have stabilized...

Page 43: ...TemProbe temperature This is due to the added resistance of the additional MultiTemp s and cable s This effect is likely to be negligible at very low temperatures but may be 0 25 F per additional Mult...

Page 44: ...hed on the display immediately prior to the display of the air flow reading The flow will be calculated using this assumed standard temperature 9 2 BACKPRESSURE COMPENSATION The air flow delivery of a...

Page 45: ...Figures 10 4 through 10 8 to determine the frame channels needed to assemble any of the standard size frames Each frame corner section utilizes an eyelet and slot arrangement which self locks into a s...

Page 46: ...a 2 x4 top as shown in Figures 10 5 and 10 6 The 1 x5 top requires an 8 5 dowel extender at the bottom of each support dowel The dowel end pins are to be inserted into the inner set of frame support c...

Page 47: ...ADM 860C 04 14 08 42 FIGURE 10 1 FRAME STORAGE FIGURE 10 2 FLOWHOOD IN CASE...

Page 48: ...ADM 860C 04 14 08 43 FIGURE 10 3 FLOWHOOD ASSEMBLY...

Page 49: ...ADM 860C 04 14 08 44 FIGURE 10 4 2X2 FRAME ASSEMBLY FIGURE 10 5 1X4 FRAME ASSEMBLY FIGURE 10 6 2X4 FRAME ASSEMBLY...

Page 50: ...ADM 860C 04 14 08 45 FIGURE 10 7 1X5 FRAME ASSEMBLY FIGURE 10 8 3X3 FRAME ASSEMBLY...

Page 51: ...h hands one on the handle and the other at any convenient point on the outside of the base Set the flaps control knob to the open position Press the READ key or the thumbswitch on the FlowHood handle...

Page 52: ...compensated and noncompensated readings are recorded separately in memory Individual readings displayed using the RCL keys will have a small sc for stored backpressure compensated readings or a small...

Page 53: ...the change from cool set up to heated working conditions A fan which was set to deliver 5000 cfm at 75 F intake air will actually deliver only about 4610 cfm of the 75 F room air when the air is heat...

Page 54: ...et size is not practical in most cases The standard 30 length top sizes may be used with equal accuracy on supply or return air outlets However small supply outlets at high jet velocity Punca diffuser...

Page 55: ...e sensor is used during flow or velocity measurements If the temperature probe is not used the meter calculates the air density using an assumed temperature of 70 F If the TemProbe sensor was not used...

Page 56: ...hermal anemometers or hot wire instruments which measure mass flow mass flux unit time Mass flow represents the number of molecules of air flowing past a given point during a given time Mass flow only...

Page 57: ...e independently of any other meter manometer or U tube To avoid pulsation or cross talk through the pressure source tubing use shutoff valves to isolate the AirData Multimeter and the reference pressu...

Page 58: ...ood and the velocity measurement using the AirData Multimeter and pitot tube This means that if the TemProbe is used it must be used for both types of measurements If the TemProbe is not used for eith...

Page 59: ...arger Any attempt to service or repair anything inside the meter will void the Warranty and may cause serious damage to the sensitive electronic components The AirData Multimeter should be returned to...

Page 60: ...al cushioning to fill the carton Do not use spray foam Spray foam can damage the meter and it is also possible to lose a meter in a chunk of spray foam Ship directly to Shortridge Instruments Inc 7855...

Page 61: ...ular charges will apply This warranty shall be effective only if use and maintenance is in accordance with our instructions and written notice of a defect is given to the Seller within such period THI...

Page 62: ...s for the density effect of variations in temperature Record the AirData Multimeter temperature and absolute barometric pressure reading at the beginning and end of each probe test range These reading...

Page 63: ...e set up with different pressures of precisely the desired test differential pressure These testers must be carefully adjusted to an equal base pressure of approximately 3 psi before adding the weight...

Page 64: ...e battery set by placing your DC voltmeter common ground probe at the left center battery clip and the other probe at the upper right corner battery clip The voltage should be between 7 2 volts and 8...

Page 65: ...ed above The temperature during storage should be less than 95 F The batteries will function best after storage if reconditioned by performing two full cycles of 24 hour charge and full discharge foll...

Page 66: ...ndle bracket 8 tubing harness MC84 1 Fitted foam lined carrying case IM 860C 1 Instruction manual ADM 860C SWE 1 1 WinWedge software sample template PR 245 K 1 Seiko portable printer with accessories...

Page 67: ...plete frame assembly all of above 1x5S 1 Cloth skirt 13 x 60 827E 4 Support dowel extenders 1x5T 1 1x5 top set complete 3 x 3 TOP 91 4 x 91 4 cm 141 4 Side channel type 1 336 4 Side channel type 6 3x3...

Page 68: ...l buttons 23 correction factors velocity 26 27 cursor control 23 data bits 19 22 data download to a computer 20 data download to printer 19 data settings 22 DENS 17 density correction 2 35 50 53 densi...

Page 69: ...P 0 13 ON OFF key 4 6 OPEN FLAPS 13 operational temperature limits 3 outlet resistance 39 output buffer size 22 OVER FLOW 13 OVER PRES 13 OVER TEMP 13 36 OVER VEL 13 Pa nnn nn 8 Pa READ 7 parity 19 22...

Page 70: ...e connection 27 traverse 27 TREND RDY 15 TREND readings 17 tubing 3 UNDER TEMP 15 36 UNITS 4 UNITS XX 15 vapor pressure 50 VelGrid 2 15 31 VelGrid air velocity 31 velocities low 29 velocity 7 12 13 16...

Reviews: