ADM-860C 04/14/08
18
of the first reading. Readings will be displayed as nn
R
nn, where nn
R
is the sequence number of the reading
and nn is the actual reading. Subsequent readings will be displayed in succession.
The reading and memory entry process will continue as long as the READ key is pressed (up to a maximum
of 100 readings). Press the STORE key at any point to view the sum and the average. The number of readings
that have been taken, along with the average of the readings will be displayed, such as 07
0
395. The first two
digits are the number of readings taken (the average of 100 readings is displayed as 00
0
nnn). Press the
STORE key again to display the sum of the readings. The number of readings in memory will not be displayed
with the sum.
Press the STORE key again to display the minimum reading. The sequence number of the reading will be
displayed along with the minimum reading, such as xx
º
n.nnnn. Press the STORE key again to display the
maximum reading. The sequence number of the reading will be displayed along with the maximum reading,
such as xx
¸
n.nnnn. The measurement sequence may be resumed by pressing the READ key, and may be
interrupted at any point by pressing the STORE key again.
The measurement mode or other conditions may not be changed while the meter is in the memory function.
Air flow readings may be saved as either a sequence of nonbackpressure compensated readings, or as a
sequence of alternating nonbackpressure compensated and backpressure compensated readings, with display
of the sum, average, minimum and maximum available for each type of reading. See Section 11.1 AIR FLOW -
NONBACKPRESSURE COMPENSATED READINGS and Section 11.2 AIR FLOW - BACKPRESSURE
COMPENSATED READINGS for more information.
The TemProbe must not be connected or disconnected during a STORE sequence. If this occurs, TEMPROBE
will be displayed followed by CHANGE. The TemProbe must be reconnected or disconnected before further
readings can be taken. The STORE mode may be exited by pressing SHIFT/CLEAR if no further readings are
needed.
If the storage capacity of 100 readings has been reached, STORE FULL will be displayed. Press the STORE
key to alternately display the average and sum, or one of the RCL keys for recall of the individual readings. The
memory must be cleared by pressing SHIFT/CLEAR before the next reading sequence may be begun.
5.3.2 RECALL
The RCL keys are used during a reading sequence to sequentially recall all readings that are in memory.
=
RCL
displays the last reading stored, and then displays readings in reverse order. RCL
<
displays the first reading
stored, and then displays readings in the order entered. These keys may be used either before or after the
STORE key has been pressed. The sequence of the reading will be displayed along with the reading, such as
07s 395. A brief press of a RCL key will advance the display one number at a time. Holding a recall key down
will fast forward or fast reverse the display through all the numbers that are in memory.
5.3.3 REPLACE READING IN MEMORY
A reading in memory may appear to be abnormal, or may have been taken at the wrong time, or with an
improperly positioned sensor. Readings stored in memory may be replaced with a new measurement. Press
one of the RCL keys to toggle through the individual readings in the stored sequence. Select a specific reading
such as: 07s 395. This reading may be replaced while it is on display by pressing the MODE key. REPL n?
will be displayed where n is the sequence number for that reading. Press the READ key. The meter will initiate
a new reading and the result will replace the value previously stored for that position in the sequence.
REPLACING will be displayed while the meter takes the new reading and stores it to memory. The new sum
and average will be calculated using the new measurement and may be displayed by pressing the STORE key.
5.3.4 ERASE READING IN MEMORY
Readings stored in memory may be erased without replacement. Press one of the RCL keys to toggle through
the individual readings in the stored sequence. Select a specific reading such as: 07s 395. This reading may
be erased while it is on display by pressing the MODE key twice. ERASE n? will be displayed where n is the
sequence number for that reading. Press the READ key. ERASING will be displayed. The erased reading will
Summary of Contents for ADM-860C
Page 38: ...ADM 860C 04 14 08 33 FIGURE 6 3 VELGRID ASSEMBLY...
Page 47: ...ADM 860C 04 14 08 42 FIGURE 10 1 FRAME STORAGE FIGURE 10 2 FLOWHOOD IN CASE...
Page 48: ...ADM 860C 04 14 08 43 FIGURE 10 3 FLOWHOOD ASSEMBLY...
Page 50: ...ADM 860C 04 14 08 45 FIGURE 10 7 1X5 FRAME ASSEMBLY FIGURE 10 8 3X3 FRAME ASSEMBLY...