9. 

Performance test 

The performance of the column can be checked by calculating the number of theoretical plates under 
the following conditions: 
1) Specimen and injection amount: 

KS-801  1.0% ethylene glycol aqueous solution  5 µL 
KS-802  2.5% ethylene glycol aqueous solution   5µL 
KS-803  0.5% ethylene glycol aqueous solution  15 µL 
KS-804  0.8% glucose aqueous solution  5 µL 
KS-805  0.4% glucose aqueous solution  5 µL 
KS-806  0.8% glucose aqueous solution  5 µL 

2) Mobile phase: 

  

distilled water 

3) Flow rate:    

 

1.0 mL/min 

4) Column temperature:  

50°C 

5) Detector:  

 

 

High-sensitivity refractive index detector such as Shodex RI 

6) Calculation formula:   

NTP = 5.54 x (Rt/W)2 

where   NTP: Number of theoretical plate, Rt: Retention time, W: Peak half width 

 

 

Summary of Contents for SUGAR KS-800 Series

Page 1: ... Shodex SUGAR KS 800 series Europe Middle East Africa Russia www shodex de Sales Office Showa Denko Europe GmbH Shodex Business Konrad Zuse Platz 3 81829 Munich Germany Technical Support info shodex de 49 0 89 93 99 62 37 Quotes and Orders order shodex de 49 0 89 93 99 62 34 ...

Page 2: ...r ion Particle size µm Exclusion limit Pullulan Plate number TP per column Size mm Shipping solvent F6378010 SUGAR KS 801 Sulfo Na 6 1 000 17 000 8 0 x 300 H2O F6378020 SUGAR KS 802 Sulfo Na 6 10 000 17 000 8 0 x 300 H2O F6378025 SUGAR KS 803 Sulfo Na 6 50 000 17 000 8 0 x 300 H2O F6378035 SUGAR KS 804 Sulfo Na 7 400 000 17 000 8 0 x 300 H2O F6378050 SUGAR KS 805 Sulfo Na 17 5 000 000 9 000 8 0 x ...

Page 3: ...n such a way that the arrow mark on the column will point to the direction of the flow Caution Do not let the air into the column during the connection 4 Heat the column to 40 50 C and start the pump Caution i Keep the column pressure below 5 MPa per column otherwise the column will be ruined ii The maximum working flow rate depends on the working temperature as follows Working temperature Maximum...

Page 4: ...3 Cap both ends of the column place it back in the same case in which it was delivered from the manufacturer and store the case in a place where the temperature does not vary more than 5 C from the ambient temperature Caution i Do not remove the end fittings of the column otherwise its performance will deteriorate or the column may be rendered unserviceable ii Do not let the in column solvent free...

Page 5: ...S 802 2 5 ethylene glycol aqueous solution 5µL KS 803 0 5 ethylene glycol aqueous solution 15 µL KS 804 0 8 glucose aqueous solution 5 µL KS 805 0 4 glucose aqueous solution 5 µL KS 806 0 8 glucose aqueous solution 5 µL 2 Mobile phase distilled water 3 Flow rate 1 0 mL min 4 Column temperature 50 C 5 Detector High sensitivity refractive index detector such as Shodex RI 6 Calculation formula NTP 5 ...

Page 6: ... following cases 3 1 If the Shodex product is not used in accordance with the operating manual 3 2 If the Shodex product is remodelled by anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not con...

Reviews: