Method 1: Apply after using the column under highly aqueous condition. 
Introduce water/acetonitrile = 20/80. 

 

Method 2: Apply to samples containing proteins. 
Introduce water/acetonitrile = 50/50 - 30/70 (containing 0.1 % TFA). 

 

 

Attention!

   

· Some proteins may precipitate in high-concentration organic solvents. 

 

· Keeping acidic solution in the column for a long time will lead to the column deterioration. 

Complete the suggested column cleaning steps all the way up to the eluent replacement 
at once and do not hold the procedure in middle way.     

 

8. Column Storage

 

Remove the column from the system after replacing the in-column solvent with the shipping solvent. Securely 
tighten the end caps and store the column at a location with stable temperature (a cool and dark space is 
recommended). Refer to section 7.3 Solvent Exchange for how to replace the eluent. 

 

 

Attention!

   

· Never allow inside of the column to dry. It can damage the column.     

 

9. Column Inspection

 

Inspection method is described in the Certificate of Analysis (CoA).   
Theoretical Plate Number (N) and Asymmetry Factor (Fas) were calculated using the below equations.   

 

(1) Theoretical Plate Number (N) 

 

 

(2) Asymmetry Factor (Fas) 

 

 

 

10. Additional Warnings 

(1) Do not remove end fittings. 

 

(2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. 

 

Refer to the Shodex™ website (https://www.shodex.com/) for product details and their applications.   
For additional assistance, contact the distributor from whom you purchased the column or contact your regional 
Shodex

 support office (https://www.shodex.com/en/support_office/list/). 

 
 

 

Summary of Contents for Silica C18M 4D

Page 1: ...check if product name and serial number written on the column package column adhesive label and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex Silica C18M series is a monomeric type ODS column filled with high purity silica gel modified with octadecyl group 2 Column Components 3 Column Specifications Product Code Prod...

Page 2: ...ication system or water from a newly opened HPLC grade distilled water bottle Use HPLC grade organic solvent whenever possible Solvents left in an opened bottle for a long time should not be used The content may have been changed absorbed moisture or has been contaminated Always use freshly prepared solvents Solvents stored for a long time may have changed their compositions and may influence elut...

Page 3: ...d deteriorates the column Note It is recommended to set the pump limiter to avoid exceeding the maximum pressure 7 3 Solvent Exchange 1 When replacing the solvent start the system at less than 0 5 mL min Recommended solvent volume to introduce at each step is 3 to 5 times of the column volume 2 Check miscibility solubility of the desired new solvent and the solvent currently filled in the column 3...

Page 4: ...ten the end caps and store the column at a location with stable temperature a cool and dark space is recommended Refer to section 7 3 Solvent Exchange for how to replace the eluent Attention Never allow inside of the column to dry It can damage the column 9 Column Inspection Inspection method is described in the Certificate of Analysis CoA Theoretical Plate Number N and Asymmetry Factor Fas were c...

Page 5: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Reviews: