Shodex RSpak DM-614 Operation Manual Download Page 4

ver.210707E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

7.3 Solvent Exchange 

(1) When replacing the solvent, start the system at less than 0.3 mL/min. Recommended solvent volume to 

introduce at each step is 3 to 5 times of the column volume. 

 

(2) Check miscibility/solubility of the desired new solvent and the solvent currently filled in the column. 

 

(3)

 

When replacing the current solvent with a solvent with low miscibility/solubility to the current solvent, first 
use a solvent that is miscible/soluble to both eluents, and then replace it with the new solvent. 

 

e.g. When replacing highly concentrated buffer solution or salt solution to water/acetonitrile, first run water 

and then replace it with the eluent. 

 

(4) When using a gradient method, changes in the eluent compositions may increase the column backpressure. 

Adjust the flow rate and column temperature so that the column backpressure remains below the usable 
maximum pressure.

 

 

7.4 Column Cleaning 

Problems in peak shapes and elution time changes or elevated column pressure etc. are often caused by 
insoluble or adsorbing components present in the eluent and reagents being deposited inside the column. These 
problems may be resolved by cleaning the column.   
If a guard column is used with the analytical column, first remove the guard column and check the performance 
of the analytical column alone. If the problem is solved, most likely the cause is from the guard column. In this 
case, clean the guard column.   
If the problem is not solved by removing the guard column, clean both guard and analytical columns. Make sure 
to clean the guard and the analytical columns separately. When washing the column, let the washing solution 
flow from the column outlet go directly into the waste container and not let the solution go through the detector. 

 

<Cleaning Method> 
(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., 

introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate. 

 

(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction. 

 

 

Method: 
Set the flow rate at 0.3 mL/min, and introduce 50 mL of methanol, acetonitrile or water.     
 

8. Column Storage

 

Remove the column from the system after replacing the in-column solvent with water. Securely tighten the end 
caps and store the column at a location with stable temperature (a cool and dark space is recommended). Refer 
to section 7.3 Solvent Exchange for how to replace the eluent. 

 

 

Attention!

   

· Never allow inside of the column to dry. It can damage the column.     

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for RSpak DM-614

Page 1: ...1 Please visually inspect the package and outside of the column for any damage 2 Please check if product name and serial number written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex RSpak DM 614 is a polymer based reversed phase chromatography column It can be used in a wide ...

Page 2: ...ances may result in deterioration of the column and or they appear as noise on the chromatograms Filter the eluent with a 0 45 μm disposable filter to prevent the problems Attention Whenever water is required use ultra pure water freshly generated by a water purification system or water from a newly opened HPLC grade distilled water bottle Use HPLC grade organic solvent whenever possible Solvents ...

Page 3: ... the end fitting and secure it with the male nut It is important that there is no extra space between the tubing and the column side of the end fitting Presence of an extra space will let the sample to spread out and may result in wide peaks 3 Set the initial flow rate at less than 0 3 mL min and start the system If using the column at an elevated temperature keep a low flow rate until the tempera...

Page 4: ...the column If a guard column is used with the analytical column first remove the guard column and check the performance of the analytical column alone If the problem is solved most likely the cause is from the guard column In this case clean the guard column If the problem is not solved by removing the guard column clean both guard and analytical columns Make sure to clean the guard and the analyt...

Page 5: ...late Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support of...

Page 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Reviews: