Shodex IC YS-50 Operation Manual Download Page 4

ver.211201E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

Method 1: Adsorption of hydrophilic compounds   

Washing Solution 

Cleaning Time 

1 H

2

O  

30 

min 

4 mM Methanesulfonic acid aq. solution/CH

3

CN = 70/30 

100 min 

3 H

2

O  

60 

min 

4 Eluent 

100 

min 

 

 

Method 2: Adsorption of hydrophobic compounds   

Washing Solution 

Cleaning Time 

1 H

2

O  

30 

min 

20 mM Methanesulfonic acid aq. solution 

  60 min 

3 H

2

O  

30 

min 

4 Eluent 

100 

min 

 
8. Column Storage

 

Remove the column from the system after replacing the in-column solvent with the shipping solvent. Securely 
tighten the end caps and store the column at a location with stable temperature (a cool and dark space is 
recommended). Refer to section 7.3 Solvent Exchange for how to replace the eluent. 

 

 

Attention!

   

· Never allow inside of the column to dry. It can damage the column.     

 

9. Column Inspection

 

Inspection method is described in the Certificate of Analysis (CoA).   
Theoretical Plate Number (N) and Asymmetry Factor (Fas) were calculated using the below equations.   

 

(1) Theoretical Plate Number (N) 

 

 

(2) Asymmetry Factor (Fas) 

 

 

 
 
 
 
 
 

Summary of Contents for IC YS-50

Page 1: ...ackage and outside of the column for any damage 2 Please check if product name and serial number written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex IC YS 50 is an ion chromatography column suitable for a simultaneous analysis of monovalent and divalent cations It is applic...

Page 2: ...d use ultra pure water freshly generated by a water purification system or water from a newly opened HPLC grade distilled water bottle Use HPLC grade acetonitrile whenever possible acetonitrile left in an opened bottle for a long time should not be used The content may have been changed absorbed moisture or has been contaminated Always use freshly prepared solvents Solvents stored for a long time ...

Page 3: ...e pump was stopped while the eluent inside the column was still hot as the eluent temperature decreases its volume also decreases This may result in creating an empty space in the column and deteriorates the column Note It is recommended to set the pump limiter to avoid exceeding the maximum pressure 7 3 Solvent Exchange When replacing the solvent start the system at less than 0 5 mL min Recommend...

Page 4: ... Storage Remove the column from the system after replacing the in column solvent with the shipping solvent Securely tighten the end caps and store the column at a location with stable temperature a cool and dark space is recommended Refer to section 7 3 Solvent Exchange for how to replace the eluent Attention Never allow inside of the column to dry It can damage the column 9 Column Inspection Insp...

Page 5: ...ct on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support office https www shodex com en support_office list ...

Page 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Reviews: