7.2 Column Installation

 

 

(1) Connect the column to LC system by following the “flow direction arrow” (

) indicated on the column name 

tag. If guard column is used, position the guard column in front (before the inlet) of the analytical column.   

 

(2) Make sure to insert the tubing all the way to the end fitting and secure it with the male nut. It is important that 

there is no extra space between the tubing and the column side of the end fitting. Presence of an extra space 
will let the sample to spread out and may result in wide peaks. 

 

 

 

 

(3)

 

Set the initial flow rate at less than 0.5 mL/min and start the system. If using the column at an elevated 

temperature, keep a low flow rate until the temperature of the column reaches to the set temperature, and 
then gradually increase the flow rate to the desired.   

 

(4)

 

When replacing the shipping solvent with the eluent, wash the column with the eluent for more than        

60 minutes.

 

 

 Caution! 

 

 

· Verify that there is no solvent leak. It may cause electronic leakage, rust, and/or chemical 

injury. 

 

 

Attention!

   

·

 

Make sure not to let air bubbles enter the column while installing the column. The air 
bubbles may damage the column.   

 

· When restarting the system after column installation or after holding the eluent flow, start 

the system at less than 0.5 mL/min. A rapid increase in pressure can damage the column.   

 

· If the column was used at an elevated temperature, lower the flow rate to less than 0.5 

mL/min at the end of analysis. Then, turn off the column oven, and let the column 
temperature return to room temperature before stopping the pump. If the pump was 
stopped while the eluent inside the column was still hot, as the eluent temperature 
decreases, its volume also decreases. This may result in creating an empty space in the 
column and deteriorates the column. 

 

       

Note  

 

 

· It is recommended to set the pump limiter to avoid exceeding the maximum pressure.   

 

7.3 Solvent Exchange 

When replacing the solvent, start the system at less than 0.5 mL/min. Recommended solvent volume to 
introduce at each step is 3 to 5 times of the column volume. 

 

 

Attention!

   

· Frequent solvent replacement deteriorates the column, and thus not recommended. 

 
7.4 Column Cleaning 

Problems in peak shapes and elution time changes or elevated column pressure etc. are often caused by 
insoluble or adsorbing components present in the eluent and reagents being deposited inside the column. These 
problems may be resolved by cleaning the column.   
If a guard column is used with the analytical column, first remove the guard column and check the performance 
of the analytical column alone. If the problem is solved, most likely the cause is from the guard column. In this 
case, clean the guard column.   
If the problem is not solved by removing the guard column, clean both guard and analytical columns. Make sure 
to clean the guard and the analytical columns separately. When washing the column, let the washing solution 
flow from the column outlet go directly into the waste container and not let the solution go through the detector. 

 

<Cleaning Method> 
(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., 

introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate. 

 

(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction 

and use the flow rate at 0.5 mL/min. 

Summary of Contents for IC YS-50

Page 1: ...ackage and outside of the column for any damage 2 Please check if product name and serial number written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex IC YS 50 is an ion chromatography column suitable for a simultaneous analysis of monovalent and divalent cations It is applic...

Page 2: ...d use ultra pure water freshly generated by a water purification system or water from a newly opened HPLC grade distilled water bottle Use HPLC grade acetonitrile whenever possible acetonitrile left in an opened bottle for a long time should not be used The content may have been changed absorbed moisture or has been contaminated Always use freshly prepared solvents Solvents stored for a long time ...

Page 3: ...e pump was stopped while the eluent inside the column was still hot as the eluent temperature decreases its volume also decreases This may result in creating an empty space in the column and deteriorates the column Note It is recommended to set the pump limiter to avoid exceeding the maximum pressure 7 3 Solvent Exchange When replacing the solvent start the system at less than 0 5 mL min Recommend...

Page 4: ... Storage Remove the column from the system after replacing the in column solvent with the shipping solvent Securely tighten the end caps and store the column at a location with stable temperature a cool and dark space is recommended Refer to section 7 3 Solvent Exchange for how to replace the eluent Attention Never allow inside of the column to dry It can damage the column 9 Column Inspection Insp...

Page 5: ...ct on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support office https www shodex com en support_office list ...

Page 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Reviews: