Shodex IC YS-50 Operation Manual Download Page 1

ver.211201E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

Operation Manual 
Shodex

 IC YS-50 

 

(Please read this manual carefully to achieve the best and consistent column performance for a long time) 

 

 

Important Handling Instructions 

 

 Caution! 

 

 

· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column 

and understand their proper handling methods to prevent potential health hazards or death 
from occurring. 

 

·

 

Please wear appropriate personal protective equipment such as lab goggles and gloves 

when handling organic solvents and acid and alkaline reagents. Avoid any direct physical 
contact to prevent chemical injuries.   

 

Before Using the Column 

(1) Please visually inspect the package and outside of the column for any damage.   

 

(2) Please check if product name and serial number written on the column package, column name tag, and     

enclosed Certificate of Analysis (CoA) are matching and correct.   

 

1. Introduction 

Thank you for purchasing the Shodex

 product. Shodex

 IC YS-50 is an ion chromatography column 

suitable for a simultaneous analysis of monovalent and divalent cations. It is applicable for both suppressor 
and non-suppressor methods.   

 

2. Column Components   

 

 
 
 
 
 
 
 
 
 
 
 
 
3. Column Specifications 

Product 

Code 

Product Name 

Column Size (mm) 

Particle 

Size (

μ

m) 

Theoretical Plate Number 

(Per Column) 

I.D. Length 

F7122000 

IC YS-50 

4.6 125  5 

 5,500 

F6700530 

IC YS-G 

4.6  

10  5 

(Guard 

Column) 

 

Base Material 

: Spherical porous particles of polyvinyl alcohol modified with carboxyl group

 

Column Housing  : SUS-316 
Screw Type 

: Internally-threaded type No.10-32 UNF 

Shipping Solvent : 10 mM Sodium Sulfate aq. Solution 
 
 
 
 
 
 

Summary of Contents for IC YS-50

Page 1: ...ackage and outside of the column for any damage 2 Please check if product name and serial number written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex IC YS 50 is an ion chromatography column suitable for a simultaneous analysis of monovalent and divalent cations It is applic...

Page 2: ...d use ultra pure water freshly generated by a water purification system or water from a newly opened HPLC grade distilled water bottle Use HPLC grade acetonitrile whenever possible acetonitrile left in an opened bottle for a long time should not be used The content may have been changed absorbed moisture or has been contaminated Always use freshly prepared solvents Solvents stored for a long time ...

Page 3: ...e pump was stopped while the eluent inside the column was still hot as the eluent temperature decreases its volume also decreases This may result in creating an empty space in the column and deteriorates the column Note It is recommended to set the pump limiter to avoid exceeding the maximum pressure 7 3 Solvent Exchange When replacing the solvent start the system at less than 0 5 mL min Recommend...

Page 4: ... Storage Remove the column from the system after replacing the in column solvent with the shipping solvent Securely tighten the end caps and store the column at a location with stable temperature a cool and dark space is recommended Refer to section 7 3 Solvent Exchange for how to replace the eluent Attention Never allow inside of the column to dry It can damage the column 9 Column Inspection Insp...

Page 5: ...ct on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support office https www shodex com en support_office list ...

Page 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Reviews: