Shodex Asahipak GF-HQ Series Operation Manual Download Page 5

ver.211228E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

7.3 Solvent Exchange 

(1) When replacing the solvent, start the system at less than 0.2 mL/min. Recommended solvent volume to 

introduce at each step is 3 to 5 times of the column volume. 

 

(2) Check miscibility/solubility of the desired new solvent and the solvent currently filled in the column. 

 

(3) When replacing the current solvent with a miscible solvent, first introduce 1:1 mixture of the current solvent 

and the new solvent, and then replace it with 100 % of the new solvent. 

 

e.g. When replacing water to THF, first run 1:1 mixture of water and THF and then introduce 100 % THF. 

 

(4) When replacing the current solvent with a solvent with low miscibility/solubility to the current solvent, first 

use a solvent that is miscible/soluble to both eluents, and then replace it with the new solvent. 

 

e.g. When replacing THF to DMF with LiBr, first follow step (3) above, and then replace 100 % DMF to DMF 

with LiBr.   

 

 

Attention!

   

· Frequent solvent replacement deteriorates the column, and thus not recommended. 

 

7.4 Column Cleaning 

Problems in peak shapes and elution timing changes or elevated column pressure etc. are often caused by 
insoluble or adsorbing components present in the eluent and reagents being deposited inside the column. These 
problems may be resolved by cleaning the column.   
If a guard column is used with the analytical column, first remove the guard column and check the performance 
of the analytical column alone. If the problem is solved, most likely the cause is from the guard column. In this 
case, clean the guard column.   
If the problem is not solved by removing the guard column, clean both guard and analytical columns. Make sure 
to clean the guard and the analytical columns separately. When washing the column, let the washing solution 
flow from the column outlet go directly into the waste container and not let the solution go through the detector. 

 

<Cleaning Method> 
(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., 

introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate. 

 

(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction 

and use the flow rate less than 0.3 mL/min. Introduce 5 to 10 times column volume of the washing solvent. 

 

 

Method 1: Adsorption of hydrophilic compounds (when using aqueous eluent) 

Introduce the eluent with a high polar organic solvent (acetonitrile or methanol) added. 

 

Method 2: Adsorption of ionic compounds 

  Introduce the eluent with a higher salt concentration.     

 

 

Attention!

   

· Keep the organic solvent and salt within the concentrations stated in 4.2 Usable Solvents. 

 

8. Column Storage

 

Remove the column from the system after replacing the in-column solvent with the shipping solvent. Securely 
tighten the end caps and store the column at a location with stable temperature (a cool and dark space is 
recommended). Refer to section 7.3 Solvent Exchange for how to replace the eluent. 

 

 

Attention!

   

· Never allow inside of the column to dry. It can damage the column.     

 
 
 
 
 
 
 
 

 

Summary of Contents for Asahipak GF-HQ Series

Page 1: ...ial number written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex Asahipak GF HQ series is polymer based size exclusion chromatography columns They work well with both aqueous and organic solvents 2 Column Components 3 Column Specifications Product Code Product Name Column Siz...

Page 2: ...ium sulfate Please keep total concentration of salts under 0 5 M Their recommended concentration ranges are 0 05 0 3 M 2 Salts such as lithium bromide can be added to DMF and DMSO Their recommended concentration ranges are 10 50 mM 3 Aqueous solutions of protein denaturants such as urea and guanidine hydrochloride can also be used However since highly concentrated solutions are usually used there ...

Page 3: ...ould not be used The content may have been changed absorbed moisture or has been contaminated Always use freshly prepared solvents Solvents stored for a long time may have changed their compositions and may influence elution patterns and or damage the column Note Use of on line degasser is recommended 6 Sample Preparation 1 If possible use the eluent for analysis to dissolve or dilute samples If t...

Page 4: ... If using the column at an elevated temperature keep a low flow rate until the temperature of the column reaches to the set temperature and then gradually increase the flow rate to the desired 4 Multiple number of columns can be connected in series When connecting multiple columns with each having different exclusion limits set the column with higher exclusion limit at the upper stream position Wh...

Page 5: ...cal column alone If the problem is solved most likely the cause is from the guard column In this case clean the guard column If the problem is not solved by removing the guard column clean both guard and analytical columns Make sure to clean the guard and the analytical columns separately When washing the column let the washing solution flow from the column outlet go directly into the waste contai...

Page 6: ...late Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support of...

Page 7: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...