Shodex Asahipak GF-HQ Series Operation Manual Download Page 4

ver.211228E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

7. Column Usage Procedure 

 

7.1 HPLC System Preparation 

Wash entire LC system prior to the column installation, including all flow-lines and sample loop by switching the 
valve, and then replace the washing solution with the eluent to be used. If desired new eluent has low 
miscibility/solubility to the eluent of previous analysis, first use the eluent that is miscible/soluble to both eluents, 
and then replace it with the desired eluent.   

 

e.g. When replacing chloroform to water, first run methanol and then replace it with water. 

 

e.g. When replacing highly concentrated buffer solution or salt solution to water/acetonitrile, first run water and 

then replace it with water/acetonitrile. 

 

 Attention! 

 

· If the eluent left in the system is not compatible with the column to be used, it may damage 

the column. 

 

 

     · A drastic change in the eluent compositions may remove substances adsorbed on the 

system and they may enter and deteriorate the column. 

 

7.2 Column Installation

 

 

(1) Connect the column to LC system by following the “flow direction arrow” (

) indicated on the column name 

tag. If guard column is used, position the guard column in front (before the inlet) of the analytical column.   

 

(2) Make sure to insert the tubing all the way to the end fitting and secure it with the male nut. It is important that 

there is no extra space between the tubing and the column side of the end fitting. Presence of an extra space 
will let the sample to spread out and may result in wide peaks.   

 

 

 

(3) Set the initial flow rate at less than 0.2 mL/min and start the system. If using the column at an elevated 

temperature, keep a low flow rate until the temperature of the column reaches to the set temperature, and 
then gradually increase the flow rate to the desired.   

 

(4) Multiple number of columns can be connected in series. When connecting multiple columns with each having 

different exclusion limits, set the column with higher exclusion limit at the upper stream position. When 
connecting GF-7M HQ, use multiple number of the same columns in series. This is because

 

GF-7M HQ is 

prepared by using several different gels with different pore sizes, designed to provide linear calibration 
curves over a set molecular weight range. If a GF-7M HQ was used together with columns with different 
pore sizes, this changes the mixing ratio of the gels and consequently it may deform the linear calibration 
curve.   

 

 Caution! 

 

 

· Verify that there is no solvent leak. It may cause electronic leakage, rust, and/or chemical 

injury. 

 

 

Attention!

   

· Make sure not to let air bubbles enter the column while installing the column. The air 

bubbles may damage the column.   

 

· When restarting the system after column installation or after holding the eluent flow, start 

the system at less than 0.2 mL/min. A rapid increase in pressure can damage the column.   

 

· If the column was used at an elevated temperature, lower the flow rate to less than      

0.2 mL/min at the end of analysis. Then, turn off the column oven, and let the column 
temperature return to room temperature before stopping the pump. If the pump was 
stopped while the eluent inside the column was still hot, as the eluent temperature 
decreases, its volume also decreases. This may result in creating an empty space in the 
column and deteriorates the column. 

 

       

Note  

 

 

· It is recommended to set the pump limiter to avoid exceeding the maximum pressure.   

 

Summary of Contents for Asahipak GF-HQ Series

Page 1: ...ial number written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex Asahipak GF HQ series is polymer based size exclusion chromatography columns They work well with both aqueous and organic solvents 2 Column Components 3 Column Specifications Product Code Product Name Column Siz...

Page 2: ...ium sulfate Please keep total concentration of salts under 0 5 M Their recommended concentration ranges are 0 05 0 3 M 2 Salts such as lithium bromide can be added to DMF and DMSO Their recommended concentration ranges are 10 50 mM 3 Aqueous solutions of protein denaturants such as urea and guanidine hydrochloride can also be used However since highly concentrated solutions are usually used there ...

Page 3: ...ould not be used The content may have been changed absorbed moisture or has been contaminated Always use freshly prepared solvents Solvents stored for a long time may have changed their compositions and may influence elution patterns and or damage the column Note Use of on line degasser is recommended 6 Sample Preparation 1 If possible use the eluent for analysis to dissolve or dilute samples If t...

Page 4: ... If using the column at an elevated temperature keep a low flow rate until the temperature of the column reaches to the set temperature and then gradually increase the flow rate to the desired 4 Multiple number of columns can be connected in series When connecting multiple columns with each having different exclusion limits set the column with higher exclusion limit at the upper stream position Wh...

Page 5: ...cal column alone If the problem is solved most likely the cause is from the guard column In this case clean the guard column If the problem is not solved by removing the guard column clean both guard and analytical columns Make sure to clean the guard and the analytical columns separately When washing the column let the washing solution flow from the column outlet go directly into the waste contai...

Page 6: ...late Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support of...

Page 7: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Reviews: