10.2 Self Diagnosis Function (ERROR)
This instrument always diagnoses itself on the items listed in the Table 10.1 below. When an error is
found in the self diagnostics results, a relevant error number is displayed.
10.2.1 Self Diagnosis Items
㩷 㩷 㩷 㩷 㩷 㩷 㩷
Table 10.1
㩷
List of Diagnosis Items
Type
Error
message
Parts
Diagnostic
Remedy
Zero point
sensor
Zero point position of servomotor
cannot be detected.
ADC
conversion end error, Initializing error
IC for clock
The clock battery is used up
䌛
Note1
䌝
WDT
Watchdog timer error
EEP WRITE
Non-volatile memory Writing error
EEP
READ1
Non-volatile memory Reading error
(Setup data)
EEP
READ2
Non-volatile memory Reading error
(Engineering data)
Ha
rd w
are e
rror
EEP
READ3
Non-volatile memory Reading error
(ADC Calibration data)
Contact our
dealer where
you purchased
the instrument
or our sales
representative.
Area error
Input value is beyond the setting range.
Time error
Clock time setting error
Setting range
error
Setting the channel (CH1) which cannot
set the sum/difference/average
operation.
Area error
Span L
㻢
Span R
Area error
Scale L
㻢
Scale R
Area
error
Zone L
㻢
Zone R or
Zone R
䋭
Zone L is less than 5mm.
Reference
channel setting
error
The reference channel range of the
sum/difference/average operation is
except Volt, TC, RTD and SCALE.
Ope
ration e
rror
All channels
skip error
Setting the all channels to Skip.
Set properly
Other
Other
Not calibrating with the specified input.
Calibrate it with
correct input.
䌛
Note1
䌝
The battery life is supposed to last approx.10 years but sometimes depends on the environment
it is used.