5. Specifications 

Measurement Range 

Temperature: 0 to 50     Humidity: 5 to 90 %RH 

Output    

Temperature Humidity 

DSW-100 

-TRH, -TRH W4

100 

/0   (Pt100), 3-wire type 

4 to 20 mA DC (Linear conversion 
corresponding to 0 to100 %RH) 
2-wire or 4-wire type (Selectable) 
Maximum allowable load: 500 

 max. 

DSW-100 

-TAH, -TAH W4 

4 to 20 mA DC (Linear conversion 
corresponding to 0 to 50 ) 
2-wire or 4-wire type (Selectable) 
Maximum allowable load: 500 

 max. 

DSW-100 
-TRHV W4 

100 

/0   (Pt100), 3-wire type 

0 to 1 V DC (Linear conversion 
corresponding to 0 to100 %RH) 
4-wire type only 
Maximum allowable load: 500 

 max. 

DSW-100 
-TAHV W4 

4 to 20 mA DC (Linear conversion 
corresponding to 0 to 50 ) 
2-wire or 4-wire type (Selectable) 
Maximum allowable load: 500 

 max. 

Power Supply Voltage 

24 V DC 10% 

Element Type 

Temperature: RTD 100 

/0   (Pt100) (IEC 60751) 

Humidity: Electrostatic capacity change-type 

Mounting 

To the outlet box (inside wall), via outlet box cover. Mounting dimension: 66.7 mm 
(JIS-C8340: 1999) (JIS: Japan Industrial Standards.) 

Material 

Flame-resistant PC resin,    Color: White 

Wiring 

Lead wire: 300 mm    Cross-section area: 0.5 mm

2

  See “3. Wiring” for details. 

External Dimensions 

W90 x H90 x D32 mm (excluding lead wire) 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Weight 

Approx. 90 g 

Performance 

Accuracy 

Temp.: DSW-100-TRH, -TRH W4, -TRHV W4: 

(0.1+0.0017|t|)  

DSW-100-TAH, -TAH W4, -TAHV W4: 

0.5  

Humidity: 

5 %RH (at 5 to 45 )  

Out of range 5 to 45 : Max. 

8 %RH 

Under the conditions of:   

Input power supply: 24 V, Load resistance: 250 

,  

Wind velocity: 0.15 m/s, Warm-up period: 60 minutes 

The output accuracy will be affected if the conditions above are 
changed.

 

Response 
characteristics 

Temp.: 63% response within 1 minute (at 1.5 m/s wind velocity)   

Humidity: Approx. 20 seconds [Time to reach 90% of the RH value 
when the RH shifts between 30        85 %RH. However, airflow 
5 l/min (0.16 m/s)] (Waterproof filter attached)   

Hysteresis: Approx. 0 %RH (S

tabilization time: 20 minutes)

   

Power Consumption 

Approx. 550 mW (2-wire type),   
Approx. 3 W (4-wire type: DSW-100-TRH W4, DSW-100-TRHV W4)   
Approx. 4 W (4-wire type: DSW-100-TAH W4, DSW-100-TAHV W4) 

Operating  
Environment 

Temperature: 0 to 50     Humidity: 5 to 90 %RH (non-condensing) 

[Caution]   
Do not use this sensor in an environment where dew condensation occurs. 
Do not use this sensor in an environment where chlorinated and sulfidizing 
gases are being generated.

 

Storage Environment Temperature: -20 to 60     Humidity: 5 to 90 %RH (non-condensing)   
Insulation Resistance 

Between Case - Output: 500 M   minimum, at 500 V DC 

Dielectric Strength   

Between Case - Output: 1.5 kV for 1 minute, 3 mA max. 

Accessories 

Sensor main unit mounting screw: M3 x 4 (4 pieces), Mounting Plate, Mounting Plate 
screw: M4 x 8 (2 pieces), Spring washer (2 pieces), Instruction manual: 1 copy 

Environmental Spec. 

RoHS directive compliant 

 

Summary of Contents for DSW-100-TH

Page 1: ...dures which may lead to dangerous conditions and cause death or serious injury if not carried out properly Caution Procedures which may lead to dangerous conditions and cause superficial to medium injury or physical damage or may degrade or damage the product if not carried out properly Warning To prevent an electrical shock or fire only Shinko or other qualified service personnel may handle the i...

Page 2: ...should be examined giving careful consideration to the following conditions This sensor is designed for indoor use only Do not install outside This instrument is intended to be used under the following environmental conditions Free air flow A minimum of dust and an absence of corrosive gases No flammable explosive gases No mechanical vibrations or shocks No exposure to direct sunlight an ambient t...

Page 3: ...downward NOTE As the humidity sensor HD S2 is polarity sensitive so be sure to insert the sensor correctly otherwise the sensor will be broken and measurement will be impossible Never disassemble the humidity sensor as it may lead to malfunction 4 Mount the main unit cover then turn the power to the unit ON 1 Fixing the Mounting Plate Screw either A or B and fix the Mounting Plate to the outlet bo...

Page 4: ... 0 1 0 0017 t DSW 100 TAH TAH W4 TAHV W4 0 5 Humidity 5 RH at 5 to 45 Out of range 5 to 45 Max 8 RH Under the conditions of Input power supply 24 V Load resistance 250 Wind velocity 0 15 m s Warm up period 60 minutes The output accuracy will be affected if the conditions above are changed Response characteristics Temp 63 response within 1 minute at 1 5 m s wind velocity Humidity Approx 20 seconds ...

Page 5: ...全上のご注意 ご ご使 使用 用前 前に に必 必ず ずお お読 読み みく くだ ださ さい い 安全上のご注意では 安全注意事項のランクを 警告 注意 として区分しています なお 注 意に記載した事項でも 状況によっては重大な結果に結びつく可能性がありますので 記載 している事柄は必ず守ってください 取扱いを誤った場合 危険な状況が起こりえて 人命や重大な傷害にかかわる事 故の起こる可能性が想定される場合 取扱いを誤った場合 危険な状況が起こりえて 中程度の傷害や軽傷を受ける可 能性が想定される場合および機器損傷の発生が想定される場合 警 告 感電および火災防止の為 弊社のサービスマン以外は本器内部に触れないでください 感電 火災事故および機器故障防止の為 部品の交換は弊社のサービスマン以外は行わないでください 安全に関するご注意 正しく安全にお使いいただくため ご使用の前には必ず本書...

Page 6: ...0 50 で急激な温度変化および氷結の可能性がないところ 湿度が 5 90 RH で 結露の可能性がないところ 塩素系 硫化系の気体が生じる可能性がないところ 大容量の電磁開閉器や 大電流の流れている電線から離れているところ 水 油および薬品またはそれらの蒸気が直接あたる恐れのないところ お問い合わせ 本器について不明な点がございましたら 大変お手数ですが本器の形名 計器番号をご確認の上 弊社営 業所へお問い合わせください なお 動作上の不具合については その内容とご使用状態の詳細を具体的にお知らせください 本 社 562 0035 大阪府箕面市船場東 2 丁目 5 番 1 号 TEL 072 727 4571 FAX 072 727 2993 URL http www shinko technos co jp 大 阪 営 業 所 562 0035 大阪府箕面市船場東 2 丁目 5 番 1...

Page 7: ...図 3 1 図 3 2 図 3 3 図 3 4 DSW 100 TRH TRH W4 TRHV W4 DSW 100 TAH TAH W4 TAHV W4 リード線の色 リード線の種類 リード線の色 リード線の種類 RED 赤 A 100 0 Pt100 RED 赤 4 20mA DC WHT 白 B WHT 白 WHT 白 B ORG 橙 4 20mA DC または 0 1V DC ORG 橙 4 20mA DC または 0 1V DC YLW 黄 YLW 黄 BRW 茶 電源 BRW 茶 電源 BLK 黒 BLK 黒 4 湿度センサ HD S2 の交換 湿度センサ HD S2 の交換手順を 以下に示します 本器の電源を OFF し 本体カバーを取り外してください 湿度センサ HD S2 をソケットから抜いてください 新しい湿度センサ HD S2 の形名を表示した銘板の面が下になるよう ...

Page 8: ...0 5 mm2 詳細は 3 結線 をご参照ください 外形寸法 W90 H90 D32 mm リード線を除く 質 量 約 90 g 性 能 精 度 温度 DSW 100 TRH TRH W4 TRHV W4 0 1 0 0017 t DSW 100 TAH TAH W4 TAHV W4 0 5 湿度 5 RH at 5 45 5 45 範囲外の場合 最大 8 RH 入力電源電圧 24 V 負荷抵抗 250 風速 0 15 m s ウォームアップ時間 60 min において この条件と異なる場合は出力精度に影響する 応答特性 温度 1 分以内 63 応答 1 5 m s の風速の場合 湿度 約 20 秒 30 85 RH を 100 とした時の 90 到達時間 ただし エア フロー5 l min 0 16 m s 防水フィルター実装済み ヒステリシス 約 0 RH 安定時間 20 分 消費電力...

Reviews: